NJFET-PMOS PREAMPLIFIER SYSTEM - AN UPGRADED VERSION

被引:7
作者
BUTTLER, W
LUTZ, G
MANFREDI, PF
RE, V
SPEZIALI, V
TOMASINI, A
机构
[1] MAX PLANCK INST PHYS & ASTROPHYS,W-8000 MUNICH 40,GERMANY
[2] UNIV PAVIA,DIPARTIMENTO ELETTR,I-27100 PAVIA,ITALY
[3] IST NAZL FIS NUCL,I-20133 MILAN,ITALY
关键词
D O I
10.1016/0168-9002(92)90739-Q
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The NJFET-PMOS monolithic preamplifier system JAMEX64 is now available in three versions: they differ in the power consumption and in the minimum gate length of the NJFETs. The most recent version employs short channel NJFETs (L = 1.6-mu-m), to open-up the possibility of extending the range of application of the system to accelerators of bunch-crossing in the 100 ns range. The paper describes the weighting function associated with the JAMEX64 analog channels and compares the results of the theoretical noise analysis with the experimental measurements of the equivalent noise charge.
引用
收藏
页码:420 / 424
页数:5
相关论文
共 6 条
[1]   A MONOLITHIC RADIATION-TOLERANT CHARGE-SENSITIVE PREAMPLIFIER [J].
BUTTLER, W ;
CITTERIO, M ;
LUTZ, G ;
MANFREDI, PF ;
RE, V ;
SPEZIALI, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 292 (02) :435-438
[2]   JFET-PMOS TECHNOLOGY IN THE DESIGN OF MONOLITHIC PREAMPLIFIER SYSTEMS FOR MULTIELECTRODE DETECTORS [J].
BUTTLER, W ;
VOGT, H ;
LUTZ, G ;
MANFREDI, PF ;
SPEZIALI, V .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1991, 38 (02) :69-76
[3]  
BUTTLER W, 1990, 4TH TOP SEM EXP APP, P394
[4]  
BUTTLER W, 1990, NUCL INSTRUM METH A, V268, P140
[5]  
GATTI E, 1986, RIV NUOVO CIMENTO, V9, P62
[6]   HIGH-DENSITY MONOLITHIC FRONT-END SYSTEMS FOR DETECTORS WITHOUT CHARGE MULTIPLICATION [J].
MANFREDI, PF ;
SPEZIALI, V .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 279 (1-2) :152-168