COMPARISON OF BREATH-HOLD T1-WEIGHTED MR SEQUENCES FOR IMAGING OF THE LIVER

被引:21
作者
SEMELKA, RC
WILLMS, AB
BROWN, MA
BROWN, ED
FINN, JP
机构
[1] SIEMENS TRAINING & DEV CTR,CARY,NC
[2] SIEMENS MED SYST INC,ISELIN,NJ
来源
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING | 1994年 / 4卷 / 06期
关键词
LIVER; MR; PULSE SEQUENCES; RAPID IMAGING;
D O I
10.1002/jmri.1880040602
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Three rapid T1-weighted gradient-echo techniques for imaging of the liver were compared: fast low-angle shot (FLASH) and section-selective (SSTF) and non-section-selective (NSTF) inversion-recovery TurboFLASH. Ten healthy volunteers were imaged at 1.5 T, with breath-hold images acquired in the transaxial and coronal planes and non-breath-hold images in the transaxial plane. Breath-hold images were evaluated quantitatively and qualitatively, and non-breath-hold images were evaluated qualitatively. FLASH images had significantly higher (P < .001) spleen-liver signal difference-to-noise ratios (SD/Ns) than NSTF and SSTF images. Liver signal-to-noise ratios (S/Ns) were significantly higher (P < .001) on FLASH images than on NSTF and SSTF images. With breath hold, FLASH images were rated as having the highest quality in the axial plane, followed by NSTF and SSTF images. In the coronal plane. NSTF images were rated as having the highest quality. For images acquired during patient respiration, NSTF images had the highest quality and showed the least degradation. The results suggest that FLASH images have the highest SD/N and S/N for liver imaging and have the highest quality in the axial plane. In patients who cannot suspend respiration, NSTF images may be least affected by breathing artifact and provide reasonable image quality.
引用
收藏
页码:759 / 765
页数:7
相关论文
共 9 条
[1]   STRATEGIES TO IMPROVE CONTRAST IN TURBOFLASH IMAGING - REORDERED PHASE ENCODING AND K-SPACE SEGMENTATION [J].
CHIEN, D ;
ATKINSON, DJ ;
EDELMAN, RR .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (01) :63-70
[2]  
CHIEN D, 1991, Magnetic Resonance Quarterly, V7, P31
[3]   MAGNETIZATION PREPARED RAPID GRADIENT-ECHO (MP-RAGE) MR IMAGING OF THE LIVER - COMPARISON WITH SPIN-ECHO IMAGING [J].
DELANGE, EE ;
MUGLER, JP ;
BERTOLINA, JA ;
GAY, SB ;
JANUS, CL ;
BROOKEMAN, JR .
MAGNETIC RESONANCE IMAGING, 1991, 9 (04) :469-476
[4]   MR-IMAGING OF THE LIVER - BREATH-HOLD T1-WEIGHTED MP-GRE COMPARED WITH CONVENTIONAL T2-WEIGHTED SE IMAGING - LESION DETECTION, LOCALIZATION, AND CHARACTERIZATION [J].
DELANGE, EE ;
MUGLER, JP ;
BOSWORTH, JE ;
DEANGELIS, GA ;
GAY, SB ;
HURT, NS ;
BERR, SS ;
ROSENBLATT, JM ;
MERICKEL, LW ;
HARRIS, EK .
RADIOLOGY, 1994, 190 (03) :727-736
[5]   SELECTIVE VERSUS NONSELECTIVE PREPARATION PULSES IN 2-DIMENSIONAL MP-RAGE IMAGING OF THE LIVER [J].
DELANGE, EE ;
MUGLER, JP ;
BERTOLINA, JA ;
BROOKEMAN, JR .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (03) :355-358
[6]   SEGMENTED TURBOFLASH - METHOD FOR BREATH-HOLD MR IMAGING OF THE LIVER WITH FLEXIBLE CONTRAST [J].
EDELMAN, RR ;
WALLNER, B ;
SINGER, A ;
ATKINSON, DJ ;
SAINI, S .
RADIOLOGY, 1990, 177 (02) :515-521
[7]   THE IMPORTANCE OF PHASE-ENCODING ORDER IN ULTRA-SHORT TR SNAPSHOT MR IMAGING [J].
HOLSINGER, AE ;
RIEDERER, SJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 16 (03) :481-488
[8]   A SIMPLE METHOD OF GENERATING VARIABLE T1 CONTRAST IMAGES USING TEMPORALLY REORDERED PHASE ENCODING [J].
NORRIS, DG ;
BOTTCHER, U ;
LEIBFRITZ, D .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (03) :483-490
[9]   T1-WEIGHTED SEQUENCES FOR MR IMAGING OF THE LIVER - COMPARISON OF 3 TECHNIQUES FOR SINGLE-BREATH, WHOLE-VOLUME ACQUISITION AT 1.0-T AND 1.5-T [J].
SEMELKA, RC ;
SIMM, FC ;
RECHT, M ;
DEIMLING, M ;
LENZ, G ;
LAUB, GA .
RADIOLOGY, 1991, 180 (03) :629-635