CHEMICALLY MODIFIED ION-SENSITIVE FIELD-EFFECT TRANSISTORS - ELIMINATION OF THE LIQUID JUNCTION POTENTIAL IN A DOUBLE SENSOR FLOW-INJECTION ANALYSIS CELL

被引:12
作者
COBBEN, PLHM
EGBERINK, RJM
BOMER, JG
SCHOUWENAAR, R
BRZOZKA, Z
BOS, M
BERGVELD, P
REINHOUDT, DN
机构
[1] UNIV TWENTE,CHEM ANAL LAB,7500 AE ENSCHEDE,NETHERLANDS
[2] UNIV TWENTE,BIOINFORMAT LAB,7500 AE ENSCHEDE,NETHERLANDS
关键词
FLOW INJECTION; ION SELECTIVE ELECTRODES; SENSORS; ION-SENSITIVE FIELD EFFECT TRANSISTORS; LEAD; POTASSIUM;
D O I
10.1016/0003-2670(93)80403-8
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow-through cell was designed that can be used for flow-injection analysis with two chemically modified ion-sensitive field-effect transistors (CHEMFETs) in close proximity. This offers the possibility of a differential measurement without influence of the liquid junction potential. The differential signal of a potassium- and a sodium-selective CHEMFET responded in a Nernstian manner and selectively towards potassium in an excess (fixed) concentration of sodium. A highly lead-selective CHEMFET was used in the same flow cell in combination with a potassium-selective CHEMFET.
引用
收藏
页码:347 / 352
页数:6
相关论文
共 18 条
[1]   THE OPERATION OF AN ISFET AS AN ELECTRONIC DEVICE [J].
BERGVELD, P .
SENSORS AND ACTUATORS, 1981, 1 (01) :17-29
[2]  
BERGVELD P, 1989, SENSOR ACTUATOR, V18, P307
[3]  
BIEDERMANN G, 1974, CHEM SCRIPTA, V5, P153
[4]   MEASUREMENTS OF PH AND PNA IN CELLS WITHOUT A LIQUID JUNCTION [J].
BOTTARI, E ;
FESTA, MR .
ANALYST, 1989, 114 (12) :1623-1626
[5]  
Brodie Leo, 1981, STARTING FORTH
[6]   CHEMICALLY MODIFIED FIELD-EFFECT TRANSISTORS - A SODIUM-ION SELECTIVE SENSOR BASED ON CALIX[4]ARENE RECEPTOR MOLECULES [J].
BRUNINK, JAJ ;
HAAK, JR ;
BOMER, JG ;
REINHOUDT, DN ;
MCKERVEY, MA ;
HARRIS, SJ .
ANALYTICA CHIMICA ACTA, 1991, 254 (1-2) :75-80
[7]   CHEMICALLY MODIFIED ION-SENSITIVE FIELD-EFFECT TRANSISTORS - APPLICATION IN FLOW-INJECTION ANALYSIS CELLS WITHOUT POLYMERIC ENCAPSULATION AND WIRE BONDING [J].
COBBEN, PLHM ;
EGBERINK, RJM ;
BOMER, JG ;
SUDHOLTER, EJR ;
BERGVELD, P ;
REINHOUDT, DN .
ANALYTICA CHIMICA ACTA, 1991, 248 (02) :307-313
[8]  
FLOGEL E, 1985, FORTH
[9]   PHOTOPOLYMERIZATION OF PLASTICIZER IN ION-SENSITIVE MEMBRANES ON SOLID-STATE SENSORS [J].
HARRISON, DJ ;
TECLEMARIAM, A ;
CUNNINGHAM, LL .
ANALYTICAL CHEMISTRY, 1989, 61 (03) :246-251