IDENTIFYING BODY-FLUID DISTRIBUTION BY MEASURING ELECTRICAL-IMPEDANCE

被引:34
作者
SCHELTINGA, MR [1 ]
JACOBS, DO [1 ]
KIMBROUGH, TD [1 ]
WILMORE, DW [1 ]
机构
[1] BRIGHAM & WOMENS HOSP,DEPT SURG,75 FRANCIS ST,BOSTON,MA 02115
关键词
D O I
10.1097/00005373-199211000-00012
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
The effects of critical illness on extracellular water (ECW) and total body water (TBW) were measured using (1) a multiple dilutional technique, and (2) whole body and regional bioelectrical impedance analysis (BIA) in a group of stable patients. Total body water and body resistance (R) were similar in patients when compared with normal healthy subjects (TBW: 45.1 +/- 4.5 vs. 46.2 +/- 3.4 L, p = 0.85; R: 518 +/- 42 vs. 500 +/- 22OMEGA, p = 0.70), and a significant relationship was present between these measurements (r = -0.87, p < 0.001). However, patients demonstrated an increase in ECW compared with controls (ECW: 18.6 +/- 1.3 vs. 14.7 +/- 1.1 L, p < 0.05). Expanded ECW values were associated with diminished electrical reactance (X(c)) values (38 +/- 6 vs. 70 +/- 4 OMEGA, p < 0.001) and these values were correlated (r = -0.67, p < 0.005). The ratio of X(c) to R determined across the body and each of the segments was significantly lower in patients compared with controls (at least p < 0.005) and this ratio measured across a leg was the most sensitive predictor of health (X(c)/ R greater-than-or-equal-to 0. 137) and disease (X(c)/R less-than-or-equal-to 0.101). Bioelectrical impedance analysis is a noninvasive and simple bedside technique that can be used to predict TBW and identify altered fluid distribution following critical illness.
引用
收藏
页码:665 / 670
页数:6
相关论文
共 30 条
  • [21] ALTERATIONS IN BODY-FLUID CONTENT CAN BE DETECTED BY BIOELECTRICAL IMPEDANCE ANALYSIS
    SCHELTINGA, MR
    JACOBS, DO
    KIMBROUGH, TD
    WILMORE, DW
    [J]. JOURNAL OF SURGICAL RESEARCH, 1991, 50 (05) : 461 - 468
  • [22] IMPEDANCE ELECTRODES POSITIONED ON PROXIMAL PORTIONS OF LIMBS QUANTIFY FLUID COMPARTMENTS IN DOGS
    SCHELTINGA, MR
    HELTON, WS
    ROUNDS, J
    JACOBS, DO
    WILMORE, DW
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1991, 70 (05) : 2039 - 2044
  • [23] THE MEASUREMENT OF TOTAL BODY WATER IN THE HUMAN SUBJECT BY DEUTERIUM OXIDE DILUTION - WITH A CONSIDERATION OF THE DYNAMICS OF DEUTERIUM DISTRIBUTION
    SCHLOERB, PR
    FRIISHANSEN, BJ
    EDELMAN, IS
    SOLOMON, AK
    MOORE, FD
    [J]. JOURNAL OF CLINICAL INVESTIGATION, 1950, 29 (10) : 1296 - 1310
  • [24] SCHOELLER DA, 1985, BODY COMPOSITION ASS, P24
  • [25] SEGAL KR, 1987, FED PROC, V46, P1334
  • [26] SPENCE JA, 1979, T AM SOC ART INT ORG, V25, P51
  • [27] Subramanyan R, 1980, J Assoc Physicians India, V28, P257
  • [28] VAISMAN N, 1987, AM J CLIN NUTR, V46, P1
  • [29] MEASUREMENT OF EXTRACELLULAR WATER VOLUME BY BROMIDE ION CHROMATOGRAPHY
    WONG, WW
    SHENG, HP
    MORKEBERG, JC
    KOSANOVICH, JL
    CLARKE, LL
    KLEIN, PD
    [J]. AMERICAN JOURNAL OF CLINICAL NUTRITION, 1989, 50 (06) : 1290 - 1294
  • [30] BIOIMPEDANCE - A NOVEL METHOD FOR THE DETERMINATION OF EXTRAVASCULAR LUNG WATER
    ZELLNER, JL
    SPINALE, FG
    CRAWFORD, FA