A NEW PACEMAKER ALGORITHM FOR CONTINUOUS CAPTURE VERIFICATION AND AUTOMATIC THRESHOLD DETERMINATION - ELIMINATION OF PACEMAKER AFTERPOTENTIAL UTILIZING A TRIPHASIC CHARGE BALANCING SYSTEM

被引:27
作者
FELD, GK
LOVE, CJ
CAMERLO, J
MARSELLA, R
机构
[1] UNIV CALIF SAN DIEGO,DEPT MED,DIV CARDIOL,SAN DIEGO,CA 92103
[2] TELECTRON PACING SYST INC,DENVER,CO
[3] OHIO STATE UNIV,DEPT MED,DIV CARDIOL,COLUMBUS,OH 43210
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1992年 / 15卷 / 02期
关键词
PACEMAKER EVOKED RESPONSE; CONTINUOUS CAPTURE VERIFICATION; AUTOMATED THRESHOLD DETERMINATION;
D O I
10.1111/j.1540-8159.1992.tb03061.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new pacemaker algorithm designed to automatically verify pacemaker capture and determine pacing threshold by detection of a stimulus evoked potential was studied in 20 patients undergoing permanent pacemaker implantation. To eliminate pacing stimulus afterpotential and detect an evoked response, a hardware feedback circuit and a software template matching algorithm were used to produce a triphasic charge-balanced pacing pulse. After charge balancing the pacing lead, a residual artifact is measured. A capture window is defined as the area integral of the first 24 msec of the evoked depolarization, and a capture threshold as one third the amplitude of the capture window. The maximum allowable residual artifact is one eighth the amplitude of the capture window. Once the stimulus afterpotential is eliminated and the evoked response detected, capture threshold is automatically and continuously determined and the algorithm adds a 0.8-V safety margin to the pacemaker output. This algorithm was run automatically and after simulated loss of capture, produced by manually decreasing pacer output below threshold, in the bipolar (13 patients) and unipolar (20 patients) pacing modes. in each patient loss of capture was immediately detected. The data were consistent (P = NS) between algorithm runs. During unipolar pacing the area integral of the first 24 msec of the evoked response was 412 +/- 137 versus 413 +/- 144 and the residual artifact 5.8 +/- 4.8 versus 8.1 +/- 7.5. The resulting ratio (signal/noise) of the two parameters was 150 +/- 141 versus 145 +/- 181. Automatically determined threshold was 0.69 +/- 0.43 V versus 0.69 +/- 0.42. During bipolar pacing the area integral of the first 24 msec of the evoked response was 266 +/- 102 versus 275 +/- 102 and the residual artifact 4.0 +/- 3.3 versus 3.9 +/- 3.0. The resulting ratio was 193 +/- 210 versus 157 +/- 170. The automatically determined threshold was 0.84 +/- 0.61 versus 0.87 +/- 0.72 V. This study has demonstrated that accurate detection of a stimulus evoked response is possible after charge balancing a permanent pacing lead, thus allowing for consistent and reliable verification of capture and automatic threshold determination. By continuously verifying capture and maintaining pacemaker output only 0.8 V above threshold, this automated system may increase patient safety and reduce pulse generator battery depletion.
引用
收藏
页码:171 / 178
页数:8
相关论文
共 12 条
[1]   MORPHOLOGY OF ENDOCARDIAL T-WAVES OF FUSION BEATS [J].
BOUTE, W ;
CALS, GLM ;
DENHEIJER, P ;
WITTKAMPF, FHM .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (11) :1693-1697
[2]   THE VENTRICULAR DEPOLARIZATION GRADIENT - EFFECTS OF EXERCISE, PACING RATE, EPINEPHRINE, AND INTRINSIC HEART-RATE CONTROL ON THE RIGHT VENTRICULAR EVOKED-RESPONSE [J].
CALLAGHAN, F ;
VOLLMANN, W ;
LIVINGSTON, A ;
BOVEJA, B ;
ABELS, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (07) :1115-1130
[3]   CHARACTERISTIC VARIATION IN EVOKED-POTENTIAL AMPLITUDE WITH CHANGES IN PACING STIMULUS STRENGTH [J].
CURTIS, AB ;
VANCE, F ;
MILLERSHIFRIN, K .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (04) :416-422
[4]   AUTOMATIC REDUCTION OF STIMULUS POLARIZATION ARTIFACT FOR ACCURATE EVALUATION OF VENTRICULAR EVOKED-RESPONSES [J].
CURTIS, AB ;
VANCE, F ;
MILLER, K .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (04) :529-537
[5]  
CURTIS AB, 1990, PACE, V13, P519
[6]   RAPID ASSESSMENT OF RATE AND ANTIARRHYTHMIC DRUG EFFECT ON THE MYOCARDIUM USING ASYMMETRIC BIPHASIC PULSE STIMULATION [J].
FURUKAWA, T ;
HERSCOVICI, H ;
DESAI, T ;
SAMPSELL, R ;
NOGAMI, A ;
MOROE, K ;
ROZANSKI, JJ ;
LISTER, JW .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (01) :52-64
[7]   STIMULATION CHARACTERISTICS OF A STEROID-ELUTING ELECTRODE COMPARED WITH 3 CONVENTIONAL ELECTRODES [J].
KLEIN, HH ;
STEINBERGER, J ;
KNAKE, W .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (02) :134-137
[8]  
Preston T A, 1974, Med Instrum, V8, P322
[9]  
RICKARDS AF, 1981, BRIT HEART J, V45, P56
[10]   INITIAL CLINICAL-EXPERIENCE WITH A RATE RESPONSIVE PACEMAKER [J].
SINGER, I ;
OLASH, J ;
BRENNAN, AF ;
KUPERSMITH, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (09) :1458-1464