A METHOD FOR IMPLEMENTING DYNAMIC PHOTON-BEAM INTENSITY MODULATION USING INDEPENDENT JAWS AND A MULTILEAF COLLIMATOR

被引:92
作者
YU, CX
SYMONS, MJ
DU, MN
MARTINEZ, AA
WONG, JW
机构
[1] William Beaumont Hospital, Royal Oak, MI
关键词
D O I
10.1088/0031-9155/40/5/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A mathematical model is derived for digitally controlled linear accelerators to deliver a desired photon intensity distribution by combining collimator motion and machine dose rate variations. It shows that, at any instant, the quotient of the machine dose rate and the speed of collimator motion is proportional to the gradient of the desired in-air photon fluence distribution. The model is applicable for both independently controlled collimator jaws and multileaf collimators and can be implemented by controlling different parameters to accommodate linear accelerators from different manufactures. For independent jaws, each pair of jaws creates photon fluence variations along the direction of the jaw movement. For multileaf collimators, where each leaf is independently controlled, any two-dimensional (2D) photon fluence distribution can be delivered. The model has been implemented for wedged isodose distributions using independent jaws, and 2D intensity modulation using a multileaf collimator. One-dimensional (1D) wedged isodose distributions are created by moving an independent jaw al constant speed while varying machine dose rate. 2D intensity modulation has been implemented using a 'dynamic stepping' scheme, which controls the leaf progression during irradiation at constant machine dose rate. With this automated delivery scheme, the beam delivery time for dynamic intensity modulation, which depends on the complexity of the desired intensity distribution, approaches that of conventional beam modifiers. This paper shows the derivation of the model, its application, and our delivery scheme. Examples of 1D dynamic wedges and 2D intensity modulations will be given to illustrate the versatility of the model the simplicity of its application, and the efficiency of beam delivery. These features make this approach practical for delivering conformal therapy treatments.
引用
收藏
页码:769 / 787
页数:19
相关论文
共 17 条
[1]   METHODS OF IMAGE-RECONSTRUCTION FROM PROJECTIONS APPLIED TO CONFORMATION RADIOTHERAPY [J].
BORTFELD, T ;
BURKELBACH, J ;
BOESECKE, R ;
SCHLEGEL, W .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (10) :1423-1434
[2]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[3]   OPTIMIZATION OF STATIONARY AND MOVING BEAM RADIATION-THERAPY TECHNIQUES [J].
BRAHME, A .
RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) :129-140
[4]   FITTING OF NORMAL TISSUE TOLERANCE DATA TO AN ANALYTIC-FUNCTION [J].
BURMAN, C ;
KUTCHER, GJ ;
EMAMI, B ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :123-135
[5]   THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION [J].
CONVERY, DJ ;
ROSENBLOOM, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1359-1374
[6]   A GENERALIZED PENCIL BEAM ALGORITHM FOR OPTIMIZATION OF RADIATION-THERAPY [J].
GUSTAFSSON, A ;
LIND, BK ;
BRAHME, A .
MEDICAL PHYSICS, 1994, 21 (03) :343-357
[7]  
KIJEWSKI P K, 1978, Medical Physics (Woodbury), V5, P426, DOI 10.1118/1.594440
[8]   HISTOGRAM REDUCTION METHOD FOR CALCULATING COMPLICATION PROBABILITIES FOR 3-DIMENSIONAL TREATMENT PLANNING EVALUATIONS [J].
KUTCHER, GJ ;
BURMAN, C ;
BREWSTER, L ;
GOITEIN, M ;
MOHAN, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :137-146
[9]  
LEAVITT DD, 1991, MED PHYS, V17, P87
[10]   CLINICALLY RELEVANT OPTIMIZATION OF 3-D CONFORMAL TREATMENTS [J].
MOHAN, R ;
MAGERAS, GS ;
BALDWIN, B ;
BREWSTER, LJ ;
KUTCHER, GJ ;
LEIBEL, S ;
BURMAN, CM ;
LING, CC ;
FUKS, Z .
MEDICAL PHYSICS, 1992, 19 (04) :933-944