IMPROVED BIOCOMPATIBILITY OF HEPARIN SURFACE-COATED VENTRICULAR ASSIST DEVICES

被引:19
作者
VONSEGESSER, LK
WEISS, BM
HANSELER, E
BISANG, B
LESKOSEK, B
VONFELTEN, A
TURINA, M
机构
[1] UNIV HOSP ZURICH, INST ANESTHESIOL, CH-8091 ZURICH, SWITZERLAND
[2] UNIV HOSP ZURICH, INST CLIN CHEM, CH-8091 ZURICH, SWITZERLAND
[3] UNIV HOSP ZURICH, DEPT MED, CH-8091 ZURICH, SWITZERLAND
关键词
BIOCOMPATIBILITY; HEPARIN SURFACE COATING; MECHANICAL CIRCULATORY SUPPORT; VENTRICULAR ASSIST DEVICE; HEPARINLESS PERFUSION;
D O I
10.1177/039139889201500509
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Heparin surface coated ventricular assist devices (VADs) and cannulas were evaluated in comparison to uncoated VADs in 10 bovine experiments (body weight 77 +/- 6 kg). All systems were primed with cristalloid solution. No systemic heparin was given. Left ventricular assist was started with a blood flow of 4.2 +/- 0.4 l/min and maintained over 6 hours. Besides hemodynamic monitoring, blood samples were taken at regular intervals for blood gas, hematological, biochemical and coagulation studies. All animals in the study group (coated) were assisted for the scheduled 6 hours without device failure. In the control group, however, total occlusion occurred in 1 VAD after 1 hour of left ventricular assist whereas the other 4 VADs remained functional throughout the protocol. Mixed venous oxygens saturation was preassist 56 +/- 12% for coated versus 63 +/- 11% for uncoated and the final value at 60 minutes after weaning was 58 +/- 16% versus 59 +/- 5% (NS). Mean hematocrit dropped from a baseline value of 33 +/- 4% for coated versus 29 +/- 8% for uncoated to 29 +/- 7% versus 30 +/- 5% (NS) after 6 hours of assist. There was no significant difference between the baseline values (5.7 +/- 3.0-mu-mol/l for coated versus 4.6 +/- 3.1-mu-mol/l for uncoated) and the 6-hour values (3.8 +/- 3.7-mu-mol/l versus 7.6 +/- 6.4-mu-mol/l) for mean plasma hemoglobine. The normalized platelet levels dropped after 10 minutes of assist to 91 +/- 21% for coated versus 94 +/- 49% for uncoated (NS) and 89 +/- 29% versus 65 +/- 44 at 6 hours (NS). The activated clotting time evolved from a baseline value of 127 +/- 12 s for coated versus 131 +/- 5 s for uncoated (NS) to 122 +/- 17 s versus 139 +/- 18 s at 60 minutes after assist (NS). Renal embolus score showed a mean level of 11 +/- 8 for coated versus 15 +/- 12 for uncoated. The VAD clot score showed a mean level of 0.2 +/- 0.4 for coated versus 2.2 +/- 1.6 for uncoated (p < 0.5). We conclude, that heparin coating of blood exposed surfaces provides significant improvement of VAD biocompatibility.
引用
收藏
页码:301 / 306
页数:6
相关论文
共 12 条
[11]   SUPERIOR HEMODYNAMICS IN LEFT HEART BYPASS WITHOUT SYSTEMIC HEPARINIZATION [J].
VONSEGESSER, LK ;
WEISS, BM ;
GALLINO, A ;
LESKOSEK, B ;
REDHA, F ;
VONFELTEN, A ;
TURINA, M .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1990, 4 (07) :384-389
[12]  
VONSEGESSER LK, 1991, T AM SOC ARTIF INT O