The implementation of a ''fast, automatic shimming technique by mapping along projections'' (FASTMAP) on a 2.1-Tesla whole-body system is described. The method provides a localized adjustment of all first- (X, Y, Z) and second-order (Z2, ZX, ZY, X2-Y2, 2XY) coils in 2 min. The time savings are achieved by mapping the magnetic field along six projections rather than from whole-imaging data sets. An analysis of noise error propagation suggests that in 64-ml volumes the residual linewidths obtained with the method are negligible when the signal-to-rms-noise ratio is above 30. The initial application of the method to localized C-13, P-31, and H-1 spectroscopy of the human brain resulted in linewidths of approximately 2 Hz for C-13 (144-ml volume), approximately 2.5 Hz tor P-31 (36-ml volume) and approximately 4 Hz for H-1 (36-ml volume) with symmetric and reproducible lineshapes.