Effects of scatter generated by beam-modifying absorbers in megavoltage photon beams

被引:19
作者
Islam, MK [1 ]
VanDyk, J [1 ]
机构
[1] ONTARIO CANC INST,DEPT CLIN PHYS,TORONTO,ON M4X 1K9,CANADA
关键词
broad beam transmission; first-scatter; beam-modifying absorbers; dose calculation; wedge factor;
D O I
10.1118/1.597650
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Transmission through a beam-modifying absorber consists of attenuated primary beam and scattered radiation generated by the absorber. The primary component of the transmitted beam is characterized by the narrow beam attenuation coefficient which depends upon the energy of the beam and type of the absorber. In addition to beam energy and absorber material, the scatter component also depends on field size, thickness and shape of the absorber, location of the absorber with respect to the source, and the point of calculation. Based upon Compton first-scatter, a method has been developed to calculate effective broad beam transmission through any arbitrarily shaped absorber with variable thickness for any points on and off the central axis. The method requires predetermined narrow beam attenuation coefficients as a function of thickness. Transmission calculations for various absorbers such as wedges and attenuators were performed for cobalt-60 and 6-MV beams and were compared with measured data. For a cobalt-60 beam, the measured transmission fraction through a 1.33-cm-thick absorber (alloy, consisting of 55% bismuth and 45% lead) for a field size of 24X24 cm(2) is 17% higher than the calculated value using a narrow beam attenuation coefficient. Also, for the same absorber, measured central axis transmission is as much as 3.6% higher compared to off-axis locations. The measured transmission fraction through a 1.33-cm absorber was found to differ by as much as 13% and 14% for Cobalt-60 and 6 MV, respectively, as the chamber-to-source distance was varied from 70 to 110 cm. The agreement between calculated and measured values is within 0.5% for both energies whereas conventional narrow beam calculations would have yielded errors of 18% and 19%, respectively. Similar agree ment was obtained when comparing calculated and measured wedge factors as a function of field size, with the maximum deviation being 0.7%. Measured scattered doses, due to an attenuator covering part of a beam, show a maximum for a thickness of approximately one mean-free path. This is also predicted by calculations with an agreement of 0.3%.
引用
收藏
页码:2075 / 2081
页数:7
相关论文
共 8 条
[1]   X-RAY SPECTRA FROM ACCELERATORS IN RANGE 2 TO 6 MEV [J].
BENTLEY, RE ;
JONES, JC ;
LILLICRAP, SC .
PHYSICS IN MEDICINE AND BIOLOGY, 1967, 12 (03) :301-+
[2]   GAMMA-RAY ABSORPTION COEFFICIENTS [J].
DAVISSON, CM ;
EVANS, RD .
REVIEWS OF MODERN PHYSICS, 1952, 24 (02) :79-107
[3]   BROAD BEAM AND NARROW BEAM ATTENUATION IN LIPOWITZS METAL [J].
ELKHATIB, EE ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1987, 14 (01) :135-139
[4]   THE EFFECT OF LEAD ATTENUATORS ON DOSE IN HOMOGENEOUS PHANTOMS [J].
ELKHATIB, EE ;
PODGORSAK, EB ;
PLA, C .
MEDICAL PHYSICS, 1986, 13 (06) :928-935
[5]   SCATTERED PHOTONS PRODUCED BY BEAM-MODIFYING FILTERS [J].
HUANG, PH ;
CHIN, LM ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1986, 13 (01) :57-63
[6]   THE CALCULATION OF TRANSMISSION THROUGH A PHOTON-BEAM ATTENUATOR USING SECTOR INTEGRATION [J].
PAPIEZ, E ;
FROESE, G .
MEDICAL PHYSICS, 1990, 17 (02) :281-286
[7]   BROAD BEAM ATTENUATION OF CO-60 GAMMA-RAYS AND 6-MV, 18-MV, AND 25-MV X-RAYS BY LEAD [J].
VANDYK, J .
MEDICAL PHYSICS, 1986, 13 (01) :105-110
[8]   ON-AXIS AND OFF-AXIS PRIMARY DOSE COMPONENT IN HIGH-ENERGY PHOTON BEAMS [J].
ZEFKILI, S ;
KAPPAS, C ;
ROSENWALD, JC .
MEDICAL PHYSICS, 1994, 21 (06) :799-808