SCANNING OPTICAL SENSOR FOR THE MEASUREMENT OF DISSOLVED-OXYGEN AND BOD

被引:36
作者
LI, XM [1 ]
RUAN, FC [1 ]
NG, WY [1 ]
WONG, KY [1 ]
机构
[1] HONG KONG POLYTECH,DEPT APPL BIOL & CHEM TECHNOL,HONG KONG,HONG KONG
关键词
BOD MEASUREMENT; DISSOLVED OXYGEN MEASUREMENT; SCANNING OPTICAL SENSOR;
D O I
10.1016/0925-4005(94)80016-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A scanning optical sensor based on the oxygen quenching of luminescence has been developed for the measurement of dissolved oxygen and five-day biological oxygen demand (BOD5). The sensing element is a silicone rubber film with embedded ruthenium(II) complex. It is attached to the inner surface of a transparent vessel in which dissolved oxygen needs to be measured. An optical scanner with an excitation light source and an optical cable connected to a light detector are used to monitor the oxygen concentration inside the vessel. The scanner provides the excitation light and senses the emitted light from the sensing element without direct physical, contact with the vessel. Therefore, one optical scanner can be used to monitor the dissolved-oxygen concentrations in many vessels in one batch. A correlation is found for linearizing the responses of luminescence quenching with the dissolved-oxygen concentration, which enables two-point calibrations of the sensors to be carried out. Effects of temperature and steam sterilization on the sensor are also studied.
引用
收藏
页码:143 / 149
页数:7
相关论文
共 15 条
[1]   DETERMINATION OF OXYGEN CONCENTRATIONS BY LUMINESCENCE QUENCHING OF A POLYMER-IMMOBILIZED TRANSITION-METAL COMPLEX [J].
BACON, JR ;
DEMAS, JN .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2780-2785
[2]   PHOTOPHYSICS AND PHOTOCHEMISTRY OF OXYGEN SENSORS BASED ON LUMINESCENT TRANSITION-METAL COMPLEXES [J].
CARRAWAY, ER ;
DEMAS, JN ;
DEGRAFF, BA ;
BACON, JR .
ANALYTICAL CHEMISTRY, 1991, 63 (04) :337-342
[3]  
Clesceri LS., 1989, STANDARD METHODS EXA, V17
[4]   OPTICAL FLUORESCENCE AND ITS APPLICATION TO AN INTRAVASCULAR BLOOD-GAS MONITORING-SYSTEM [J].
GEHRICH, JL ;
LUBBERS, DW ;
OPITZ, N ;
HANSMANN, DR ;
MILLER, WW ;
TUSA, JK ;
YAFUSO, M .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (02) :117-132
[5]  
Hitchman ML, 1978, MEASUREMENT DISSOLVE
[6]   LUMINESCENT DICYANOPLATINUM(II) COMPLEXES AS SENSORS FOR THE OPTICAL MEASUREMENT OF OXYGEN CONCENTRATIONS [J].
LEE, WWS ;
WONG, KY ;
LI, XM .
ANALYTICAL CHEMISTRY, 1993, 65 (03) :255-258
[7]   LUMINESCENT PLATINUM COMPLEX IN SOLID FILMS FOR OPTICAL SENSING OF OXYGEN [J].
LI, XM ;
WONG, KY .
ANALYTICA CHIMICA ACTA, 1992, 262 (01) :27-32
[8]   OPTICAL CHARACTERISTICS OF A RUTHENIUM(II) COMPLEX IMMOBILIZED IN A SILICONE-RUBBER FILM FOR OXYGEN MEASUREMENT [J].
LI, XM ;
RUAN, FC ;
WONG, KY .
ANALYST, 1993, 118 (03) :289-292
[9]   MECHANISM OF QUENCHING OF EMISSION OF SUBSTITUTED POLYPYRIDINERUTHENIUM(II) COMPLEXES BY IRON(III), CHROMIUM(III), AND EUROPIUM(III) IONS [J].
LIN, CT ;
BOTTCHER, W ;
CHOU, M ;
CREUTZ, C ;
SUTIN, N .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1976, 98 (21) :6536-6544
[10]   OXYGEN SENSOR FOR THE TEMPERATURE-RANGE 300-K TO 500-K BASED ON FLUORESCENCE QUENCHING OF INDICATOR-TREATED SILICONE-RUBBER MEMBRANES [J].
OPITZ, N ;
GRAF, HJ ;
LUBBERS, DW .
SENSORS AND ACTUATORS, 1988, 13 (02) :159-163