WEAR OF POLYETHYLENE ACETABULAR CUPS AGAINST ALUMINA FEMORAL HEADS - 5 PROSTHESES COMPARED IN A HIP SIMULATOR FOR 35 MILLION WALKING CYCLES

被引:30
作者
SAIKKO, VO
机构
[1] Helsinki University of Technology, Laboratory of Machine Design, SF-02150, Espoo
来源
ACTA ORTHOPAEDICA SCANDINAVICA | 1993年 / 64卷 / 05期
基金
芬兰科学院;
关键词
D O I
10.3109/17453679308993680
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
5 ultra-high molecular weight polyethylene acetabular cups articulated against alumina femoral heads for 35 million walking cycles in a hip joint simulator designed for wear tests of total hip prostheses. The specimens were from Protek, Biomet, Link, Howmedica and Thackray. In the Howmedica specimen, the wear was zero, and in Link and Thackray practically negligible. In Biomet, however, the wear was noteworthy and in Protek disastrous after 20 million cycles because in these 2 prostheses the head was attached to a titanium-alloy stem by taper-fit: titanium-alloy particles that were removed from the taper because of micromotion between the head and stem were entrapped between the head and cup, adhering to the head and making it rough, which led to severe abrasive wear of the Protek cup. It was worn through at 26 million cycles, the total wear being 3,170 mg. In Biomet, Link and Thackray, the total wear was 124, 5.3 and 17.6 mg, respectively. Polyethylene wear particles may lead to adverse tissue reactions and eventual loosening of the implant. The results indicate that by the use of alumina heads, polyethylene wear can be eliminated, but this advantage may be lost if the head is attached to a titanium-alloy taper.
引用
收藏
页码:507 / 512
页数:6
相关论文
共 12 条
[1]  
Andrisano A O, 1990, Proc Inst Mech Eng H, V204, P157, DOI 10.1243/PIME_PROC_1990_204_250_02
[2]  
[Anonymous], 1992, SOCIOLOGICA, V7
[3]  
Clarke IC, 1981, ENG MED, V10, P189
[4]  
COLLIER JP, 1991, CLIN ORTHOP RELAT R, V271, P305
[5]  
DAVIDSON JA, 1987, J BIOMED MATER RES-A, V21, P261
[6]  
Fessler H, 1989, Proc Inst Mech Eng H, V203, P15, DOI 10.1243/PIME_PROC_1989_203_004_01
[7]   CORROSION OF MODULAR HIP PROSTHESES [J].
MATHIESEN, EB ;
LINDGREN, JU ;
BLOMGREN, GGA ;
REINHOLT, FP .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (04) :569-575
[8]  
OONISHI H, 1992, J LONG-TERM EFF MED, V2, P37
[9]  
Saikko V, 1992, Proc Inst Mech Eng H, V206, P195, DOI 10.1243/PIME_PROC_1992_206_291_02
[10]   WEAR OF THE POLYETHYLENE ACETABULAR CUP - METALLIC AND CERAMIC HEADS COMPARED IN A HIP SIMULATOR [J].
SAIKKO, VO ;
PAAVOLAINEN, PO ;
SLATIS, P .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (04) :391-402