COUPLING OF HPLC AND ICP-AES FOR SPECIATION

被引:63
作者
LABORDA, F [1 ]
DELOOSVOLLEBREGT, MTC [1 ]
DEGALAN, L [1 ]
机构
[1] UNILEVER RES LABS VLAARDINGEN,3130 AC VLAARDINGEN,NETHERLANDS
关键词
D O I
10.1016/0584-8547(91)80104-B
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The cross-flow nebulizer and the thermospray vaporizer can both be used as an interface for coupling high performance liquid chromatography with inductively coupled plasma-atomic emission spectrometry. The dispersion and the peak asymmetry are acceptable for both. With the thermospray system the sensitivity is threefold higher and the fraction of methanol in methanol/water mixtures introduced into the plasma can be as high as 75%, compared with 25% for the cross-flow nebulizer. Initial results obtained for the speciation of selenium in (CH3)3Se+, SeO3(2-) and SeO4(2-) are promising.
引用
收藏
页码:1089 / 1098
页数:10
相关论文
共 31 条
[21]   ANALYTICAL PERFORMANCE OF A THERMOSPRAY SAMPLE INTRODUCTION SYSTEM FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
PENG, RZ ;
TIGGELMAN, JJ ;
DELOOSVOLLEBREGT, MTC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (1-2) :189-199
[22]   THERMOSPRAY ENHANCED INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROSCOPY DETECTION FOR LIQUID-CHROMATOGRAPHY [J].
ROYCHOWDHURY, SB ;
KOROPCHAK, JA .
ANALYTICAL CHEMISTRY, 1990, 62 (05) :484-489
[23]   EVALUATION OF SPRAY CHAMBERS FOR USE IN INDUCTIVELY COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
SCHUTYSER, P ;
JANSSENS, E .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1979, 34 (11-1) :443-449
[24]   CHARACTERIZATION OF AEROSOLS GENERATED BY THERMOSPRAY NEBULIZATION FOR ATOMIC SPECTROSCOPY [J].
SCHWARTZ, SA ;
MEYER, GA .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (12) :1287-1298
[25]   DESIGN OF COLUMN CONNECTION TUBES FOR LIQUID CHROMATOGRAPHY [J].
SCOTT, RPW ;
KUCERA, P .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1971, 9 (11) :641-&
[26]   MEASUREMENT OF AEROSOL TRANSPORT EFFICIENCY IN ATOMIC SPECTROMETRY [J].
SMITH, DD ;
BROWNER, RF .
ANALYTICAL CHEMISTRY, 1982, 54 (03) :533-537
[27]  
Sternberg J., 1966, ADV CHROMATOGR, P205
[28]   USE OF A THERMOSPRAY NEBULIZER AS A SAMPLE INTRODUCTION SYSTEM FOR INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
VERMEIREN, KA ;
TAYLOR, PDP ;
DAMS, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (04) :383-387
[29]   THERMOSPRAY LIQUID CHROMATOGRAPH MASS-SPECTROMETER INTERFACE WITH DIRECT ELECTRICAL HEATING OF THE CAPILLARY [J].
VESTAL, ML ;
FERGUSSON, GJ .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2373-2378
[30]   HIGH-POWER INDUCTIVELY COUPLED PLASMA TORCH AND IMPEDANCE MATCHING NETWORK [J].
WINDSOR, DL ;
HEINE, DR ;
DENTON, MB .
APPLIED SPECTROSCOPY, 1979, 33 (01) :56-58