CLINICAL PREDICTORS OF THE DEFIBRILLATION THRESHOLD WITH THE UNIPOLAR IMPLANTABLE DEFIBRILLATION SYSTEM

被引:66
作者
RAITT, MH [1 ]
JOHNSON, G [1 ]
DOLACK, GL [1 ]
POOLE, JE [1 ]
KUDENCHUK, PJ [1 ]
BARDY, GH [1 ]
机构
[1] UNIV WASHINGTON, DEPT MED, DIV CARDIOL, SEATTLE, WA USA
基金
美国国家卫生研究院;
关键词
D O I
10.1016/0735-1097(95)00093-J
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. The purpose of this study was to determine the relation between clinical variables and the defibrillation threshold by using a standardized testing protocol and a uniform implantable defibrillator system. Background. Past studies have not revealed useful correlations between clinical variables and the energy required to terminate ventricular fibrillation. Most of these studies did not use a uniform implantable defibrillator system or a standardized protocol to measure the defibrillation threshold and, thus, did not control for the influence of these technical influences. We postulated that defibrillator and defibrillation threshold measurement-based sources of variability overshadowed important clinical predictors. Methods. The defibrillation threshold was measured by using a standardized protocol in 101 consecutive patients. We used a transvenous unipolar pectoral defibrillation system that employed a single endocardial right ventricular defibrillation coil as the anode and the shell of an 80-cm(3) pulse generator as the cathode to deliver a 65% tilt biphasic pulse. Results. Several clinical variables were found to be significantly associated with the defibrillation threshold: patient gender, height, weight, body surface area, heart rate at rest, QRS and corrected QT (QTc) intervals, left ventricular mass and several measures of heart and chest size by chest roentgenogram. None of these variables had a correlation coefficient >0.45 with the defibrillation threshold. On multivariate analysis, left ventricular mass and heart rate at rest were the only independent predictors of the defibrillation threshold and explained only 25% of the observed variability. Conclusions. Despite the use of a uniform transvenous defibrillation system and a standardized protocol to measure the defibrillation threshold, no clinically relevant correlation was found between clinical variables and the defibrillation threshold. The defibrillation threshold is probably a function of a complex interaction of anatomic, physiologic and cellular variables that are not adequately represented by easily obtainable clinical information. It is probably not possible to predict defibrillation outcome from standard clinical variables.
引用
收藏
页码:1576 / 1583
页数:8
相关论文
共 28 条
[1]   A SIMPLIFIED, SINGLE-LEAD UNIPOLAR TRANSVENOUS CARDIOVERSION-DEFIBRILLATION SYSTEM [J].
BARDY, GH ;
JOHNSON, G ;
POOLE, JE ;
DOLACK, GL ;
KUDENCHUK, PJ ;
KELSO, D ;
MITCHELL, R ;
MEHRA, R ;
HOFER, B .
CIRCULATION, 1993, 88 (02) :543-547
[2]   EVALUATION OF ELECTRODE POLARITY ON DEFIBRILLATION EFFICACY [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
GREENE, HL .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (07) :433-437
[3]   CLINICAL-EXPERIENCE WITH A TIERED-THERAPY, MULTIPROGRAMMABLE ANTIARRHYTHMIA DEVICE [J].
BARDY, GH ;
TROUTMAN, C ;
POOLE, JE ;
KUDENCHUK, PJ ;
DOLACK, GL ;
JOHNSON, G ;
HOFER, B .
CIRCULATION, 1992, 85 (05) :1689-1698
[4]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[5]   IMPLANTABLE TRANSVENOUS CARDIOVERTER-DEFIBRILLATORS [J].
BARDY, GH ;
HOFER, B ;
JOHNSON, G ;
KUDENCHUK, PJ ;
POOLE, JE ;
DOLACK, GL ;
GLEVA, M ;
MITCHELL, R ;
KELSO, D .
CIRCULATION, 1993, 87 (04) :1152-1168
[6]   ELECTRODE SYSTEM INFLUENCE ON BIPHASIC WAVE-FORM DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
TROUTMAN, C ;
JOHNSON, G ;
MEHRA, R ;
POOLE, JE ;
DOLACK, GL ;
KUDENCHUK, PJ ;
GARTMAN, DM .
CIRCULATION, 1991, 84 (02) :665-671
[7]   PROSPECTIVE COMPARISON OF SEQUENTIAL PULSE AND SINGLE PULSE DEFIBRILLATION WITH USE OF 2 DIFFERENT CLINICALLY AVAILABLE SYSTEMS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (01) :165-171
[8]  
BHANDARI AK, 1992, CIRCULATION, V86, P790
[9]   DETERMINANTS OF SUCCESSFUL NONTHORACOTOMY CARDIOVERTER-DEFIBRILLATOR IMPLANTATION - EXPERIENCE IN 101 PATIENTS USING 2 DIFFERENT LEAD SYSTEMS [J].
BROOKS, R ;
GARAN, H ;
TORCHIANA, D ;
VLAHAKES, GJ ;
JACKSON, G ;
NEWELL, J ;
MCGOVERN, BA ;
RUSKIN, JN .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (07) :1835-1842
[10]   Successful Implantation of Cardioverter-Defibrillator Systems in Patients With Elevated Defibrillation Thresholds [J].
Brooks, Ross ;
Torchiana, David ;
Vlahakes, Gus J. ;
Ruskin, Jeremy N. ;
McGovern, Brian A. ;
Garan, Hasan .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 22 (02) :569-574