MECHANICAL LEFT-VENTRICULAR UNLOADING DURING HIGH-RISK CORONARY ANGIOPLASTY - FIRST USE OF A NEW PERCUTANEOUS TRANSVALVULAR LEFT-VENTRICULAR ASSIST DEVICE

被引:20
作者
SCHOLZ, KH
FIGULLA, HR
SCHWEDA, F
SMALLING, RW
HELLIGE, G
KREUZER, H
ABOULHOSN, W
WAMPLER, RK
机构
[1] UNIV GOTTINGEN,DEPT ANAESTHESIOL,D-37075 GOTTINGEN,GERMANY
[2] UNIV TEXAS,SCH MED,DIV CARDIOL,HOUSTON,TX
[3] JOHNSON & JOHNSON INTERVENT SYST CO,RANCHO CORDOVA,CA
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1994年 / 31卷 / 01期
关键词
PERCUTANEOUS HEMOPUMP(TM); INTRACORPOREAL CIRCULATORY SUPPORT; SUPPORTED CORONARY ANGIOPLASTY; MYOCARDIAL PROTECTION;
D O I
10.1002/ccd.1810310113
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new catheter mounted, transvalvular left ventricular assist device has been designed for percutaneous transfemoral access. The device, the Hemopump(TM) [14 French (Fr.) outer diameter], is based on a mixed flow rotary pump and is capable of flow rates of 1.5-2.2 1/min. The pump is inserted using a specialized 16 Fr. femoral introducer sheath. The first application of the percutaneous Hemopump(TM) in man was performed in two patients with hemodynamic compromise during high risk coronary angioplasty. In these patients, Hemopump(TM) support resulted in hemodynamic stabilization (increase in aortic pressure from 60/42 to 87/61 and from 80/60 to 100/70 mm Hg, respectively) and marked left ventricular unloading (decrease in pulmonary capillary wedge pressure from 25 to 10 and from 14 to 10 mm Hg) during balloon inflation. In both patients, percutaneous transluminal coronary angioplasty (PTCA) could be accomplished successfully. Using the system for periods of about 2 hr in each patient, we observed no vascular, hemorrhagic, or embolic complications. In both patients, only a minor increase in both plasma free hemoglobin and lactate dehydrogenase levels was noted. Our preliminary experiences suggest that the percutaneous Hemopump(TM) is safe and effective and may be a powerful alternative to other devices used for supported angioplasty. (C) 1994 Wiley-Liss, Inc.
引用
收藏
页码:61 / 69
页数:9
相关论文
共 13 条
[1]   PERCUTANEOUS INTRA-AORTIC BALLOON INSERTION [J].
BREGMAN, D ;
NICHOLS, AB ;
WEISS, MB ;
POWERS, ER ;
MARTIN, EC ;
CASARELLA, WJ .
AMERICAN JOURNAL OF CARDIOLOGY, 1980, 46 (02) :261-264
[2]   MYOCARDIAL SUPPORT AND PROTECTION DURING REGIONAL MYOCARDIAL-ISCHEMIA USING THE HEMOPUMP ASSIST DEVICE [J].
HERING, JP ;
SCHRODER, T ;
UHLIG, P ;
SCHOLZ, KH ;
TEBBE, U ;
KREUZER, H ;
HELLIGE, G .
THORACIC AND CARDIOVASCULAR SURGEON, 1991, 39 (05) :257-262
[3]  
HERING JP, 1992, CORONARY ARTERY DIS, V3, P419
[4]   EFFECT OF PERIPHERAL CARDIOPULMONARY BYPASS ON LEFT-VENTRICULAR SIZE, AFTERLOAD AND MYOCARDIAL-FUNCTION DURING ELECTIVE SUPPORTED CORONARY ANGIOPLASTY [J].
PAVLIDES, GS ;
HAUSER, AM ;
STACK, RK ;
DUDLETS, PI ;
GRINES, C ;
TIMMIS, GC ;
ONEILL, WW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (02) :499-505
[5]   PERCUTANEOUS INITIATION OF CARDIOPULMONARY BYPASS [J].
PHILLIPS, SJ ;
BALLENTINE, B ;
SLONINE, D ;
HALL, J ;
VANDEHAAR, J ;
KONGTAHWORN, C ;
ZEFF, RH ;
SKINNER, JR ;
RECKMO, K ;
GRAY, D .
ANNALS OF THORACIC SURGERY, 1983, 36 (02) :223-225
[6]  
Scholz K. H., 1993, European Heart Journal, V14, P475
[7]   PROTECTIVE EFFECTS OF THE HEMOPUMP LEFT-VENTRICULAR ASSIST DEVICE IN EXPERIMENTAL CARDIOGENIC-SHOCK [J].
SCHOLZ, KH ;
HERING, JP ;
SCHRODER, T ;
UHLIG, P ;
KREUZER, H ;
TEBBE, U ;
FERRARI, M ;
HELLIGE, G .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (04) :209-214
[8]  
SCHOLZ KH, IN PRESS AM HEART J
[9]   IMPROVED REGIONAL MYOCARDIAL BLOOD-FLOW, LEFT-VENTRICULAR UNLOADING, AND INFARCT SALVAGE USING AN AXIAL-FLOW, TRANSVALVULAR LEFT-VENTRICULAR ASSIST DEVICE - A COMPARISON WITH INTRAAORTIC BALLOON COUNTERPULSATION AND REPERFUSION ALONE IN A CANINE INFARCTION MODEL [J].
SMALLING, RW ;
CASSIDY, DB ;
BARRETT, R ;
LACHTERMAN, B ;
FELLI, P ;
AMIRIAN, J .
CIRCULATION, 1992, 85 (03) :1152-1159
[10]   INITIAL REPORT OF THE NATIONAL-REGISTRY-OF-ELECTIVE-CARDIOPULMONARY BYPASS SUPPORTED CORONARY ANGIOPLASTY [J].
VOGEL, RA ;
SHAWL, F ;
TOMMASO, C ;
ONEILL, W ;
OVERLIE, P ;
OTOOLE, J ;
VANDORMAEL, M ;
TOPOL, E ;
TABARI, KK ;
VOGEL, J ;
SMITH, S ;
FREEDMANN, R ;
WHITE, C ;
GEORGE, B ;
TEIRSTEIN, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (01) :23-29