OPTICAL SENSING OF PH AND PCO2 USING PHASE-MODULATION FLUOROMETRY AND RESONANCE ENERGY-TRANSFER

被引:39
作者
LAKOWICZ, JR
SZMACINSKI, H
KARAKELLE, M
机构
[1] Center for Fluorescence Spectroscopy, Department of Biological Chemistry, University of Maryland, Baltimore, MD 21201
基金
美国国家科学基金会; 美国国家卫生研究院;
关键词
FLUOROMETRY; DONOR ACCEPTOR COMPLEXES; ENERGY TRANSFER; PCO2; PH;
D O I
10.1016/0003-2670(93)80567-5
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have developed optical sensors for pH and/or pCO2 based on phase shift and modulation measurements of time-resolved fluorescence energy transfer. pH-insensitive donors and pH-sensitive acceptors were encased in CO2-permeable silicon membranes. The extent of energy transfer depends on the pH-dependent absorption spectra of the acceptor. The measurements were made insensitive to inner-filter effects of the total intensity by using the phase-modulation method to determine the donor decay times. The general applicability of this method was demonstrated using three donor-acceptor pairs, all of which can be excited with an inexpensive green 543-nm He-Ne laser.
引用
收藏
页码:179 / 186
页数:8
相关论文
共 33 条
  • [11] SINGLE FIBER OPTIC FLUORESCENCE PH PROBE
    FUH, MRS
    BURGESS, LW
    HIRSCHFELD, T
    CHRISTIAN, GD
    WANG, F
    [J]. ANALYST, 1987, 112 (08) : 1159 - 1163
  • [12] SENSITIVITY ENHANCEMENT OF FLUORESCENT PH INDICATORS BY INNER FILTER EFFECTS
    GABOR, G
    WALT, DR
    [J]. ANALYTICAL CHEMISTRY, 1991, 63 (08) : 793 - 796
  • [13] GRATTON E, 1983, BIOPHYS J, V44, P665
  • [14] END-TO-END DISTANCE DISTRIBUTIONS OF FLEXIBLE MOLECULES FROM STEADY-STATE FLUORESCENCE ENERGY-TRANSFER AND QUENCHING-INDUCED CHANGES IN THE FORSTER DISTANCE
    GRYCZYNSKI, I
    WICZK, W
    JOHNSON, ML
    LAKOWICZ, JR
    [J]. CHEMICAL PHYSICS LETTERS, 1988, 145 (05) : 439 - 446
  • [15] PHYSIOLOGICAL PH FIBEROPTIC CHEMICAL SENSOR BASED ON ENERGY-TRANSFER
    JORDAN, DM
    WALT, DR
    MILANOVICH, FP
    [J]. ANALYTICAL CHEMISTRY, 1987, 59 (03) : 437 - 439
  • [16] KAWSKI A, 1982, NATURFORSCH A, V37, P64
  • [17] A 10-GHZ FREQUENCY-DOMAIN FLUOROMETER
    LACZKO, G
    GRYCZYNSKI, I
    GRYCZYNSKI, Z
    WICZK, W
    MALAK, H
    LAKOWICZ, JR
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (09) : 2331 - 2337
  • [18] 2-GHZ FREQUENCY-DOMAIN FLUOROMETER
    LAKOWICZ, JR
    LACZKO, G
    GRYCZYNSKI, I
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1986, 57 (10) : 2499 - 2506
  • [19] CONSTRUCTION AND PERFORMANCE OF A VARIABLE-FREQUENCY PHASE-MODULATION FLUOROMETER
    LAKOWICZ, JR
    MALIWAL, BP
    [J]. BIOPHYSICAL CHEMISTRY, 1985, 21 (01) : 61 - 78
  • [20] INFLUENCE OF DIFFUSION ON EXCITATION-ENERGY TRANSFER IN SOLUTIONS BY GIGAHERTZ HARMONIC CONTENT FREQUENCY-DOMAIN FLUOROMETRY
    LAKOWICZ, JR
    SZMACINSKI, H
    GRYCZYNSKI, I
    WICZK, W
    JOHNSON, ML
    [J]. JOURNAL OF PHYSICAL CHEMISTRY, 1990, 94 (22) : 8413 - 8416