SHAPING THE SIGNAL RESPONSE DURING THE APPROACH TO STEADY-STATE IN 3-DIMENSIONAL MAGNETIZATION-PREPARED RAPID GRADIENT-ECHO IMAGING USING VARIABLE FLIP ANGLES

被引:63
作者
MUGLER, JP [1 ]
EPSTEIN, FH [1 ]
BROOKEMAN, JR [1 ]
机构
[1] UNIV VIRGINIA,HLTH SCI CTR,DEPT BIOMED ENGN,CHARLOTTESVILLE,VA 22903
关键词
D O I
10.1002/mrm.1910280202
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A theoretical algorithm for shaping the signal response during the approach to steady state in three‐dimensional magnetization‐prepared rapid gradient‐echo (3D MP‐RAGE) pulse sequences has been developed and implemented. This algorithm derives the flip angle series required to produce specifically chosen time evolutions of the signal intensities during the data acquisition segment of 3D MP‐RAGE sequences. Theoretical predictions for the cases of unshaped, uniform, and mono‐exponential decay signal responses were quantitatively validated with a doped‐water phantom on a 1.5‐T whole‐body imager and in all cases there was excellent agreement between the theoretical and experimental values. The effects of RF inhomogeneities and eddy currents on the signal response shaping were also investigated. To demonstrate the potential utility of the technique, the signal response shaping algorithm was applied to a Tl‐weighted 3D MP‐RAGE sequence to derive the acquisition flip angle series which theoretically yields the maximum white matter/gray matter signal difference (WGSD) consistent with the chosen response shape. Images obtained from a healthy volunteer using this variable flip angle sequence were compared with 3D RF‐spoiled steady‐state gradient‐echo images obtained in the same total imaging time. The 3D MP‐RAGE images demonstrated a 41% increase in the WGSD‐to‐noise ratio. These initial very promising results indicate that with further refinement to eliminate some intensity artifacts, the variable flip angle 3D MP‐RAGE technique may, with respect to certain image properties, provide considerable improvements over currently available 3D gradient‐echo imaging techniques.© 1992 Academic Press,Inc. Copyright © 1992 Wiley‐Liss, Inc., A Wiley Company
引用
收藏
页码:165 / 185
页数:21
相关论文
共 41 条
[21]   T1-WEIGHTED SNAPSHOT GRADIENT-ECHO MR IMAGING OF THE ABDOMEN [J].
HOLSINGERBAMPTON, AE ;
RIEDERER, SJ ;
CAMPEAU, NG ;
EHMAN, RL ;
JOHNSON, CD .
RADIOLOGY, 1991, 181 (01) :25-32
[22]   MULTIPLE FIELD-STRENGTH INVIVO T1 AND T2 FOR CEREBROSPINAL-FLUID PROTONS [J].
HOPKINS, AL ;
YEUNG, HN ;
BRATTON, CB .
MAGNETIC RESONANCE IN MEDICINE, 1986, 3 (02) :303-311
[23]   APPROACH TO EQUILIBRIUM IN SNAPSHOT IMAGING [J].
JONES, RA ;
RINCK, PA .
MAGNETIC RESONANCE IMAGING, 1990, 8 (06) :797-803
[24]  
KIRSCH JE, 1990, 9TH SOC MAGN RES MED, P537
[25]   RAPID 3-DIMENSIONAL T1-WEIGHTED MR IMAGING WITH THE MP-RAGE SEQUENCE [J].
MUGLER, JP ;
BROOKEMAN, JR .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (05) :561-567
[26]   3-DIMENSIONAL MAGNETIZATION-PREPARED RAPID GRADIENT-ECHO IMAGING (3DMP-RAGE) [J].
MUGLER, JP ;
BROOKEMAN, JR .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (01) :152-157
[27]   T2-WEIGHTED 3-DIMENSIONAL MP-RAGE MR IMAGING [J].
MUGLER, JP ;
SPRAGGINS, TA ;
BROOKEMAN, JR .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1991, 1 (06) :731-737
[28]  
MUGLER JP, 1990, MAGN RESON IMAGIN S1, V8, P115
[29]  
MUGLER JP, 1990, 9TH SOC MAGN RES MED, P1310
[30]  
MUGLER JP, 1991, 10TH SOC MAGN RES ME, P847