CLINICAL THERMOLUMINESCENCE DOSIMETRY - HOW DO EXPECTATIONS AND RESULTS COMPARE

被引:22
作者
KRON, T [1 ]
SCHNEIDER, M [1 ]
MURRAY, A [1 ]
MAMEGHAN, H [1 ]
机构
[1] PRINCE WALES HOSP, DEPT RADIAT ONCOL, RANDWICK, NSW 2031, AUSTRALIA
关键词
THERMOLUMINESCENCE DOSIMETRY; TLD; INVIVO DOSIMETRY; RADIOTHERAPY TREATMENT VERIFICATION;
D O I
10.1016/0167-8140(93)90097-R
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Thermoluminescence dosimetry (TLD) for radiotherapy treatment verification is performed in the Prince of Wales Hospital in Sydney for a wide range of applications: (A) to determine the dose in difficult treatment geometries, (B) to record the dose to critical organs, and (C) to monitor special treatments such as total body irradiation (TBI). TLD measurements were performed with the aim to investigate cases where dose prediction is difficult and not as part of a routine verification procedure. We reviewed 1058 reports of TLD performed during the treatment of 502 patients between 1986 and 1991 to evaluate how the TLD results compare with the dose determined by the treatment plan. Reasons for possible discrepancies should be identified. In 19% of all investigated cases a discrepancy of more than 10% was found between expected and measured doses. The discrepancies could be divided into three groups: (1) errors made in the TLD determination or evaluation, such as placement errors of the TLD chips (21% of all discrepancies); (2) mistakes made during the patient set-up, such as insufficient shielding or inadequate patient immobilisation (30%); (3) inadequate treatment planning and dose calculation procedure, such as wrong inverse square law corrections or errors due to limitations of the two-dimensional treatment planning system used (41% of all). In 8% of all discrepancies the reason remained unclear. A number of changes to treatment plans and modalities (e.g. changed scrotal shield, modified bolus) were introduced due to TLD results. The increasing number of TLD requests per year attests to the value of TLD as a treatment verification method in clinical practice.
引用
收藏
页码:151 / 161
页数:11
相关论文
共 52 条
[11]   PERIPHERAL DOSE TO THE TESTES - THE DESIGN AND CLINICAL USE OF A PRACTICAL AND EFFECTIVE GONADAL SHIELD [J].
FRAASS, BA ;
KINSELLA, TJ ;
HARRINGTON, FS ;
GLATSTEIN, E .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (03) :609-615
[12]   ROUTINE USE OF THERMOLUMINESCENCE DOSIMETRY FOR RADIATION-THERAPY [J].
GOODEN, DS ;
BRICKNER, TJ .
RADIOLOGY, 1972, 102 (03) :685-&
[13]   ASSESSMENT OF SKIN DOSE AND ITS RELATION TO COSMESIS IN THE CONSERVATIVE TREATMENT OF EARLY BREAST-CANCER [J].
HABIBOLLAHI, F ;
MAYLES, HMO ;
MAYLES, WPM ;
WINTER, PJ ;
TONG, D ;
FENTIMAN, IS ;
CHAUDARY, MA ;
HAYWARD, JL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :291-296
[14]   ENTRANCE AND EXIT TL-DOSIMETRY IN THE CONSERVATIVE TREATMENT OF BREAST-CANCER - A PILOT-STUDY FOR THE EORTC-RADIOTHERAPY COOPERATIVE GROUP [J].
HAMERS, HP ;
JOHANSSON, KA ;
VENSELAAR, JLM ;
DEBROUWER, P ;
HANSSON, U ;
MOUDI, C .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) :280-284
[15]   ABUTMENT OF HIGH-ENERGY ELECTRON FIELDS [J].
HARMS, WB ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 20 (04) :853-858
[16]   THE EYE IN ACUTE-LEUKEMIA .1. DOSIMETRIC ANALYSIS IN CRANIAL RADIATION PROPHYLAXIS [J].
HARNETT, AN ;
HIRST, A ;
PLOWMAN, PN .
RADIOTHERAPY AND ONCOLOGY, 1987, 10 (03) :195-202
[17]   COMPARISON OF ENTRANCE AND EXIT DOSE MEASUREMENTS USING IONIZATION CHAMBERS AND SILICON DIODES [J].
HEUKELOM, S ;
LANSON, JH ;
MIJNHEER, BJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (01) :47-59
[18]   INVIVO DOSIMETRY DURING TANGENTIAL BREAST TREATMENT [J].
HEUKELOM, S ;
LANSON, JH ;
VANTIENHOVEN, G ;
MIJNHEER, BJ .
RADIOTHERAPY AND ONCOLOGY, 1991, 22 (04) :269-279
[19]   ELECTRON-BEAM DOSE CALCULATIONS [J].
HOGSTROM, KR ;
MILLS, MD ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03) :445-459
[20]  
HUFTON AP, 1984, PRACTICAL ASPECTS TH