ANALYTICAL PERFORMANCE EVALUATION OF A 40.68 MHZ INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER

被引:12
作者
PARK, CJ
LEE, KW
机构
[1] Inorganic Analytical Laboratory, Korea Standards Research Institute, Taedok Science Town, Taejon
关键词
INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; 40.68 MHZ PLASMA; OPTIMIZATION AND ANALYTICAL PERFORMANCE; DETECTION LIMITS; BACKGROUND SPECTRUM;
D O I
10.1039/ja9910600431
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The instrumentation of a new 40.68 MHz inductively coupled plasma (ICP) mass spectrometer is described. An optimization study has been carried out for the instrument. Of the many operating parameters, the carrier argon flow rate is the most sensitive and the optimum carrier argon flow rate is found to be much lower than the literature values for a 27.12 MHz ICP mass spectrometer. The analytical performance of the instrument, i.e., the detection limits, background mass spectrum and formation of oxides and doubly charged ions is also evaluated.
引用
收藏
页码:431 / 437
页数:7
相关论文
共 19 条
[1]   SPECTROMETRIC STUDY ON RADIOFREQUENCY INDUCED PLASMA .1. ANALYTICAL PERFORMANCE [J].
ABDALLAH, MH ;
DIEMIASZONEK, R ;
JAROSZ, J ;
MERMET, JM ;
ROBIN, J ;
TRASSY, C .
ANALYTICA CHIMICA ACTA, 1976, 84 (02) :271-282
[2]   COMPUTER-SIMULATION OF RF INDUCTION-HEATED ARGON PLASMA DISCHARGES AT ATMOSPHERIC-PRESSURE FOR SPECTROCHEMICAL ANALYSIS .1. PRELIMINARY INVESTIGATIONS [J].
BARNES, RM ;
SCHLEICHER, RG .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1975, B 30 (04) :109-134
[3]   APPLICATIONS OF INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY TO THE PRODUCTION CONTROL OF AEROSPACE AND NUCLEAR-MATERIALS [J].
BECK, GL ;
FARMER, OT .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (06) :771-773
[4]   DETERMINATION OF URANIUM IN ENVIRONMENTAL-SAMPLES USING INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
BOOMER, DW ;
POWELL, MJ .
ANALYTICAL CHEMISTRY, 1987, 59 (23) :2810-2813
[5]   THE DETERMINATION OF TRACE-ELEMENTS IN GEOCHEMICAL-EXPLORATION SAMPLES BY ICP-MS [J].
DATE, AR ;
HUTCHISON, D .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1986, 41 (1-2) :175-181
[6]   MEASUREMENTS OF TOTAL LEAD CONCENTRATIONS AND OF LEAD ISOTOPE RATIOS IN WHOLE-BLOOD BY USE OF INDUCTIVELY COUPLED PLASMA SOURCE-MASS SPECTROMETRY [J].
DELVES, HT ;
CAMPBELL, MJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1988, 3 (02) :343-348
[7]   OXIDE AND DOUBLY CHARGED ION RESPONSE OF A COMMERCIAL INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY INSTRUMENT [J].
GRAY, AL ;
WILLIAMS, JG .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (01) :81-82
[8]   DETERMINATION OF TUNGSTEN AND MOLYBDENUM AT LOW-LEVELS IN GEOLOGICAL-MATERIALS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY [J].
HALL, GEM ;
PARK, CJ ;
PELCHAT, JC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1987, 2 (02) :189-196
[9]  
HEUMANN KG, 1982, TRENDS ANAL CHEM, V1, P357
[10]   THE EFFECT OF PLASMA OPERATING PARAMETERS ON ANALYTE SIGNALS IN INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY [J].
HORLICK, G ;
TAN, SH ;
VAUGHAN, MA ;
ROSE, CA .