BACTERIAL AND VIRAL REMOVAL EFFICIENCY, HEAT AND MOISTURE EXCHANGE PROPERTIES OF 4 FILTRATION DEVICES

被引:33
作者
VANDENBROUCKEGRAULS, CMJE
TEEUW, KB
BALLEMANS, K
LAVOOIJ, C
CORNELISSE, PB
VERHOEF, J
机构
[1] UNIV UTRECHT HOSP, INFECT DIS LAB, 3508 GA UTRECHT, NETHERLANDS
[2] EYKMAN WINKLER INST MED MICROBIOL, UTRECHT, NETHERLANDS
[3] UNIV UTRECHT HOSP, DEPT BIOMED ENGN, UTRECHT, NETHERLANDS
关键词
AIR FILTRATION; HEAT EXCHANGER; MOISTURE EXCHANGER; MECHANICAL VENTILATION; NOSOCOMIAL INFECTION;
D O I
10.1016/0195-6701(95)90292-9
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Four devices used for filtration of microorganisms and/or for heating and moistening the ventilated air during mechanical ventilation were evaluated. This evaluation included measurement of bacterial and viral removal efficiency, heat and moisture exchange properties, dead space and air flow resistance. The devices included: Pall BB50T and DAR Sterivent (filtration devices); DAR Hygrobac and Gibeck Humid-vent [heat and moisture exchangers (HMEs)]. The two devices which are primarily conceived as filters had the highest bacterial and viral removal efficiency (titre reduction of 10(5)-10(6) for bacteria and of 10(4)-10(5) for viruses), while removal efficiencies of the HME devices were lower: titre reduction of 10(4) for bacteria and 10(1)-10(3) for viruses. As expected, heat and moisture output of HMEs was better than that of filters. In mechanical ventilation, dead space and air flow resistance are important properties of devices, which might disturb efficient ventilation. There were only minor differences in dead space and air flow resistance. Resistance to airflow in the HMEs was increased by nebulization of medication (mesna) unlike that of the filters.
引用
收藏
页码:45 / 56
页数:12
相关论文
共 12 条
[1]   BACTERIAL-CONTAMINATION AND FREQUENCY OF CHANGING VENTILATOR CIRCUITRY [J].
CADWALLADER, HL ;
BRADLEY, CR ;
AYLIFFE, GAJ .
JOURNAL OF HOSPITAL INFECTION, 1990, 15 (01) :65-72
[2]  
du Moulin G C, 1977, Anesthesiology, V47, P353, DOI 10.1097/00000542-197710000-00006
[3]  
Duberstein R, 1978, J Parenter Drug Assoc, V32, P192
[4]   NOSOCOMIAL RESPIRATORY-INFECTIONS WITH GRAM-NEGATIVE BACILLI - SIGNIFICANCE OF COLONIZATION OF RESPIRATORY TRACT [J].
JOHANSON, WG ;
SANFORD, JP ;
THOMAS, GD ;
PIERCE, AK .
ANNALS OF INTERNAL MEDICINE, 1972, 77 (05) :701-+
[5]   BACTERIAL RETENTION PROPERTIES OF HEAT AND MOISTURE EXCHANGE FILTERS [J].
LEE, MG ;
FORD, JL ;
HUNT, PB ;
IRELAND, DS ;
SWANSON, PW .
BRITISH JOURNAL OF ANAESTHESIA, 1992, 69 (05) :522-525
[6]   BACTERIAL-CONTAMINATION AND THE EFFECT OF FILTERS IN ANESTHETIC CIRCUITS IN A SIMULATED PATIENT MODEL [J].
LEIJTEN, DTM ;
REJGER, VS ;
MOUTON, RP .
JOURNAL OF HOSPITAL INFECTION, 1992, 21 (01) :51-60
[7]   HEAT AND MOISTURE EXCHANGERS AND VAPORIZING HUMIDIFIERS IN THE INTENSIVE-CARE UNIT [J].
MARTIN, C ;
PERRIN, G ;
GEVAUDAN, MJ ;
SAUX, P ;
GOUIN, F .
CHEST, 1990, 97 (01) :144-149
[8]   HEAT AND MOISTURE EXCHANGERS WITH BACTERIAL FILTERS - A LABORATORY EVALUATION [J].
MEBIUS, C .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1992, 36 (06) :572-576
[9]   RISK OF CROSS-INFECTION FROM INHALATION ANAESTHETIC EQUIPMENT [J].
PANDIT, SK ;
MEHTA, S ;
AGARWAL, SC .
BRITISH JOURNAL OF ANAESTHESIA, 1967, 39 (11) :838-&
[10]   MECHANICAL VENTILATORS AS VEHICLES OF INFECTION [J].
PEREA, EJ ;
CRIADO, A ;
MORENO, M ;
AVELLO, F .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1975, 19 (03) :180-186