THE EFFECT OF HIGH-PERFORMANCE GRADIENTS ON FAST GRADIENT-ECHO IMAGING

被引:40
作者
REEDER, SB [1 ]
MCVEIGH, ER [1 ]
机构
[1] JOHNS HOPKINS UNIV,SCH MED,DEPT BIOMED ENGN & RADIOL,MED IMAGING LAB,BALTIMORE,MD 21205
关键词
OPTIMIZATION; MRI; SNR; GRADIENT COILS;
D O I
10.1002/mrm.1910320510
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effect of gradient system performance on segmented k-space gradient echo imaging is presented. Three cases were investigated. First, an ideal system that has infinite slew rates and unlimited maximum gradient strengths was considered. Second, a ''high speed'' imaging system (2.3 (G/cm), 23 (G/cm)/ms) was considered. These two cases were compared with a ''conventional'' imaging system (1 (G/cm), 1.67 (G/cm)/ms). It was found that substantial increases in SNR can be achieved (approximate to 45%) by using high speed versus a conventional gradient system, for a TR of 6 ms. For trapezoidal gradient waveforms, there exists an optimum maximum gradient strength for a given slew rate, and any increase in gradient strength above this optimum will not be utilized by an optimized sequence. These studies have shown that increasing TR without decreasing the bandwidth is not a good way to increase SNR for constant scan time.
引用
收藏
页码:612 / 621
页数:10
相关论文
共 24 条
[1]   CINEANGIOGRAPHY OF THE HEART IN A SINGLE BREATH HOLD WITH A SEGMENTED TURBOFLASH SEQUENCE [J].
ATKINSON, DJ ;
EDELMAN, RR .
RADIOLOGY, 1991, 178 (02) :357-360
[2]   ECHO-PLANAR IMAGING OF THE LIVER WITH A STANDARD MR-IMAGING SYSTEM [J].
BUTTS, K ;
RIEDERER, SJ ;
EHMAN, RL ;
FELMLEE, JP ;
GRIMM, RC .
RADIOLOGY, 1993, 189 (01) :259-264
[3]  
CONOLLY SM, 1987, J MAGN RESON, V78, P440
[4]   ELIMINATION OF TRANSVERSE COHERENCES IN FLASH MRI [J].
CRAWLEY, AP ;
WOOD, ML ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 8 (03) :248-260
[5]  
EDELSTEIN WA, 1986, MAGN RESON MED, V3
[6]  
Ernst R. R., 1987, PRINCIPLES MAGNETIC
[7]   APPLICATION OF FOURIER TRANSFORM SPECTROSCOPY TO MAGNETIC RESONANCE [J].
ERNST, RR ;
ANDERSON, WA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1966, 37 (01) :93-+
[8]   ANALYSIS OF T2 LIMITATIONS AND OFF-RESONANCE EFFECTS ON SPATIAL-RESOLUTION AND ARTIFACTS IN ECHO-PLANAR IMAGING [J].
FARZANEH, F ;
RIEDERER, SJ ;
PELC, NJ .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :123-139
[9]   PROJECTION RECONSTRUCTION TECHNIQUES FOR REDUCTION OF MOTION EFFECTS IN MRI [J].
GLOVER, GH ;
PAULY, JM .
MAGNETIC RESONANCE IN MEDICINE, 1992, 28 (02) :275-289
[10]   THE STEADY-STATE SIGNALS IN SHORT-REPETITION-TIME SEQUENCES [J].
GYNGELL, ML .