REAL-TIME, IN-VIVO MEASUREMENT OF RADIATION-DOSE DURING RADIOIMMUNOTHERAPY IN MICE USING A MINIATURE MOSFET DOSIMETER PROBE

被引:12
作者
GLADSTONE, DJ
CHIN, LM
机构
[1] Department of Radiation Oncology, Joint Center for Radiation Therapy, Harvard Medical School, Boston
关键词
D O I
10.2307/3579012
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
This report presents the first real-time measurement of absorbed radiation dose during radioimmunotherapy in mice. Dose rate and total dose at the center of the tumor were measured after administration of Y-90-labeled antibodies using a miniature metal oxide semiconductor field-effect transistor radiation dosimeter probe which was inserted into the center of the tumor volume. Continuous real-time measurements were made for as long as 23 h after injection of the radiolabeled antibodies. Comparison of the real-time dose-rate measurements with estimates based on the MIRD formalism indicates good agreement. The real-time measurements are further compared to measurements made in a second experiment in which groups of mice were sacrificed at individual times after injection of the same radiolabeled antibodies. The real-time measurements agree well with the measurements in excised tumors. The real-time measurements have greater time resolution and are much more efficient than traditional uptake measurements. (C) 1995 by Radiation Research Society
引用
收藏
页码:330 / 335
页数:6
相关论文
共 17 条
[1]   ABSORBED DOSES FOR INTERNAL RADIOTHERAPY FROM 22 BETA-EMITTING RADIONUCLIDES - BETA-DOSIMETRY OF SMALL SPHERES [J].
BARDIES, M ;
CHATAL, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (06) :961-981
[2]  
BEAUMIER PL, 1985, J NUCL MED, V26, P1172
[3]  
BERGER MJ, 1971, J NUCL MED S5, V12, P7
[4]   CMOS HARDNESS PREDICTION FOR LOW-DOSE-RATE ENVIRONMENTS [J].
DERBENWICK, GF ;
SANDER, HH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1977, 24 (06) :2244-2247
[5]   A MINIATURE MOSFET RADIATION DOSIMETER PROBE [J].
GLADSTONE, DJ ;
LU, XQ ;
HUMM, JL ;
BOWMAN, HF ;
CHI, LM .
MEDICAL PHYSICS, 1994, 21 (11) :1721-1728
[6]   AUTOMATED DATA-COLLECTION AND ANALYSIS SYSTEM FOR MOSFET RADIATION DETECTORS [J].
GLADSTONE, DJ ;
CHIN, LM .
MEDICAL PHYSICS, 1991, 18 (03) :542-548
[7]  
GRIFFITH MH, 1988, J NUCL MED, V29, P1795
[8]   THE MECHANISMS OF SMALL INSTABILITIES IN IRRADIATED MOS-TRANSISTORS [J].
HOLMESSIEDLE, A ;
ADAMS, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (06) :4135-4140
[9]   MINIATURE RADIATION DOSIMETER FOR INVIVO RADIATION MEASUREMENTS [J].
HUGHES, RC ;
HUFFMAN, D ;
SNELLING, JV ;
ZIPPERIAN, TE ;
RICCO, AJ ;
KELSEY, CA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (05) :963-967
[10]  
HUMM JL, 1994, J NUCL MED, V35, P1217