SENSITIVITY ENHANCEMENT AND 2ND-DIMENSIONAL INFORMATION FROM SOLID-PHASE EXTRACTION-CAPILLARY ELECTROPHORESIS OF ENTIRE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY FRACTIONS

被引:72
作者
STRAUSBAUCH, MA
MADDEN, BJ
WETTSTEIN, PJ
LANDERS, JP
机构
[1] MAYO CLIN & MAYO FDN,DEPT SURG,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,DEPT IMMUNOL,ROCHESTER,MN 55905
[3] MAYO CLIN & MAYO FDN,DEPT BIOCHEM & MOLEC BIOL,ROCHESTER,MN 55905
[4] MAYO CLIN & MAYO FDN,CLIN CAPILLARY ELECTROPHORESIS FACIL,ROCHESTER,MN 55905
关键词
SOLID PHASE EXTRACTION CAPILLARY ELECTROPHORESIS; ONLINE SAMPLE CONCENTRATION; SOLID PHASE; PEPTIDES;
D O I
10.1002/elps.1150160189
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A novel separation technique is demonstrated for peptide analysis of entire 100 mu L high-performance liquid chromatography (HPLC) fractions in a single, low pH, capillary zone electrophoresis (CZE) separation. The method employs a hybrid capillary consisting of a removable inlet section containing a reversephase packing and a standard CZE bare-fused silica separation capillary. The packed tip allows sample to be concentrated at flow rates greater than 20 mu L/min and released in a controlled manner for CZE separation. Separations are comparable to standard CZE with minor modifications in selectivity. Using a mixture of model peptides, the hybrid capillary is shown to effect enhancement in sensitivity of greater than or equal to 100-fold. Five fractions from the HPLC separation of peptides obtained from the digestion of bovine serum albumin (BSA) were analyzed by the solid phase extraction-capillary electrophoresis (SPE-CE) method which was shown to provide rapid second-dimensional information. Practical concentration limits of detection for peptides are calculated to be 1-10 ng/mL with this technology.
引用
收藏
页码:541 / 548
页数:8
相关论文
共 25 条
[11]  
JONES WR, 1991, AM LAB MAY, P40
[12]   ZONE ELECTROPHORESIS IN OPEN-TUBULAR GLASS-CAPILLARIES [J].
JORGENSON, JW ;
LUKACS, KD .
ANALYTICAL CHEMISTRY, 1981, 53 (08) :1298-1302
[13]   ISOTACHOPHORETIC ELECTRODESORPTION OF PROTEINS FROM AN AFFINITY ADSORBENT ON A MICROSCALE [J].
KASICKA, V ;
PRUSIK, Z .
JOURNAL OF CHROMATOGRAPHY, 1983, 273 (01) :117-128
[14]   HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS OF GLYCOPROTEINS - THE USE OF MODIFIERS OF ELECTROOSMOTIC FLOW FOR ANALYSIS OF MICROHETEROGENEITY [J].
LANDERS, JP ;
ODA, RP ;
MADDEN, BJ ;
SPELSBERG, TC .
ANALYTICAL BIOCHEMISTRY, 1992, 205 (01) :115-124
[15]  
LANDERS JP, 1995, CLIN CHEM, V41, P395
[16]   CAPILLARY ELECTROPHORESIS OF AMINO-SUGARS WITH LASER-INDUCED FLUORESCENCE DETECTION [J].
LIU, JP ;
SHIROTA, O ;
NOVOTNY, M .
ANALYTICAL CHEMISTRY, 1991, 63 (05) :413-417
[17]   ULTRASENSITIVE FLUOROMETRIC DETECTION OF CARBOHYDRATES AS DERIVATIVES IN MIXTURES SEPARATED BY CAPILLARY ELECTROPHORESIS [J].
LIU, JP ;
SHIROTA, O ;
WIESLER, D ;
NOVOTNY, M .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1991, 88 (06) :2302-2306
[18]  
MAZZEO JR, 1991, BIOTECHNIQUES, V10, P638
[19]   HIGH-PERFORMANCE ZONE ELECTROPHORESIS [J].
MIKKERS, FEP ;
EVERAERTS, FM ;
VERHEGGEN, TPEM .
JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB) :11-20
[20]   MULTIPLE-BUFFER-ADDITIVE STRATEGIES FOR ENHANCED CAPILLARY ELECTROPHORETIC SEPARATION OF PEPTIDES [J].
ODA, RP ;
MADDEN, BJ ;
MORRIS, JC ;
SPELSBERG, TC ;
LANDERS, JP .
JOURNAL OF CHROMATOGRAPHY A, 1994, 680 (01) :341-351