Accumulating resampling (modulation) in comprehensive two-dimensional capillary GC (GC x GC)

被引:36
作者
Blumberg, Leonid M. [1 ]
机构
[1] Fast GC Consulting, Wilmington, DE 19801 USA
关键词
Accumulating resampling; Optimal sampling density; Resampling; Sampling density;
D O I
10.1002/jssc.200800424
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
During each sampling period, an accumulating resampler (modulator) in comprehensive 2-D chromatography accumulates all eluite from the first-dimension column and reinjects the whole or a portion of the accumulated material into the second-dimension column. The detrimental effect of the resampling on peak capacity of a 2-D separation comes from the broadening of the peaks along the first-dimension due to the resampling itself and due to the subsequent peak reconstruction. Sampling density (rho(S)) of resampling is the number of sampling periods per standard deviation of a peak at the outlet of the first-dimension column. It is shown that a simple formula describes the peak broadening as a function of rho(S) at any (even practically too low or too high) rho(S), for the peaks of any (not necessarily Gaussian) shape, for a wide class of peak reconstruction techniques, and for any 2-D separation (GC x GC, LC x LC, etc.). In capillary GC x GC, optimal rho(S) (rho(S,Opt)) depends on the type of the peak reconstruction and on the degree of the gas decompression along the second-dimension column. When reconstructing using linear interpolation, rho(S,Opt) = 0.7 at large and rho(S,Opt) = 0.5 at small gas decompression. The choice of exact optimal conditions is not critical. Thus, two-fold departure of actual rho(S) from rho(S,Opt) in either direction (under- or oversampling) causes only 10% drop in the net peak capacity of GC x GC. The quantitative effect of a much greater undersampling is also evaluated.
引用
收藏
页码:3358 / 3365
页数:8
相关论文
共 19 条
[11]  
GUIOCHON G, 1969, ADV CHROMATOGR, P179
[12]   Calculating optimal modulation periods to maximize the peak capacity in two-dimensional HPLC [J].
Horie, Kanta ;
Kimura, Hiroshi ;
Ikegami, Tohru ;
Iwatsuka, Akira ;
Saad, Nabil ;
Fiehn, Oliver ;
Tanaka, Nobuo .
ANALYTICAL CHEMISTRY, 2007, 79 (10) :3764-3770
[13]  
Jayant N. S., 1984, DIGITAL CODING WAVEF
[14]   Modulation ratio in comprehensive two-dimensional gas chromatography [J].
Khummueng, Weeraya ;
Harynuk, James ;
Marriott, Philip J. .
ANALYTICAL CHEMISTRY, 2006, 78 (13) :4578-4587
[15]   A systematic analysis of a flow-switching modulator for comprehensive two-dimensional gas chromatography [J].
LaClair, RW ;
Bueno, PA ;
Seeley, JV .
JOURNAL OF SEPARATION SCIENCE, 2004, 27 (5-6) :389-396
[16]   COMPREHENSIVE 2-DIMENSIONAL GAS-CHROMATOGRAPHY USING AN ON-COLUMN THERMAL MODULATOR INTERFACE [J].
LIU, ZY ;
PHILLIPS, JB .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1991, 29 (06) :227-231
[17]   Effect of sampling rate on resolution in comprehensive two-dimensional liquid chromatography [J].
Murphy, RE ;
Schure, MR ;
Foley, JP .
ANALYTICAL CHEMISTRY, 1998, 70 (08) :1585-1594
[18]   Theoretical study of incomplete sampling of the first dimension in comprehensive two-dimensional chromatography [J].
Seeley, JV .
JOURNAL OF CHROMATOGRAPHY A, 2002, 962 (1-2) :21-27
[19]  
TRETTER SA, 1976, INTRO DISCRETE TIME