Electroosmotic Flow-Based Pump for Liquid Chromatography on a Planar Microchip

被引:47
作者
Borowsky, Joseph F. [1 ]
Giordano, Braden C. [2 ]
Lu, Qin [1 ]
Terray, Alex [1 ]
Collins, Greg E. [1 ]
机构
[1] USN, Res Lab, Washington, DC 20375 USA
[2] Nova Res Inc, Alexandria, VA 22308 USA
关键词
D O I
10.1021/ac801497r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An electroosmotic flow (EOF)-based pump, integrated with a sol-gel stationary phase located in the electric field-free region of a microchip, enabled the separation of six nitroaromatic and nitramine explosives and their degradation products via liquid chromatography (LC). The integrated pump and LC system were fabricated within a single quartz substrate. The pump region consisted of a straight channel (3.0 cm x 230 pm x 100 pm) packed lwith 5-mu m porous silica beads. The sol-gel stationary phase was derived from a precursor mixture of methyltrimethoxy- and phenethyltzimethoxysilanes and was synthesized in the downstream, field-free region of the microchip, resulting in a stationary-phase monolith with dimensions of 2.6 cm x 230 mu m x 100 mu m. Fluid dynamic design considerations are discussed, especially as they relate to integrating the EOF pump with the LC system. Pump and separation performance, as characterized by flow rate measurements, injection, elution, separation, and detection, point to a viable analytical chemistry platform that encompasses all of the benefits expected of portable, laboratory-on-chip systems, including reduced sample requirements and small packaging.
引用
收藏
页码:8287 / 8292
页数:6
相关论文
共 26 条
[1]  
Arnold DW, 2000, ABSTR PAP AM CHEM S, V219, pU97
[2]   High pressure electroosmotic pump based on a packed bed planar microchip [J].
Borowsky, Joseph ;
Lu, Qin ;
Collins, Greg E. .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 131 (01) :333-339
[3]  
Breadmore MC, 2002, ELECTROPHORESIS, V23, P3487, DOI 10.1002/1522-2683(200210)23:20<3487::AID-ELPS3487>3.0.CO
[4]  
2-5
[5]   Study of an electroosmotic pump for liquid delivery and its application in capillary column liquid chromatography [J].
Chen, LX ;
Ma, JP ;
Guan, YF .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1028 (02) :219-226
[6]   Electroosmosis- and pressure-driven chromatography in chips using continuous beds [J].
Ericson, C ;
Holm, J ;
Ericson, T ;
Hjertén, S .
ANALYTICAL CHEMISTRY, 2000, 72 (01) :81-87
[7]   Electrically actuated, pressure-driven liquid chromatography separations in microfabricated devices [J].
Fuentes, Hernan V. ;
Woolley, Adam T. .
LAB ON A CHIP, 2007, 7 (11) :1524-1531
[8]   Micellar electrokinetic chromatography and capillary electrochromatography of nitroaromatic explosives in seawater [J].
Giordano, BC ;
Copper, CL ;
Collins, GE .
ELECTROPHORESIS, 2006, 27 (04) :778-786
[9]   Microchip-based CEC of nitroaromatic and nitramine explosives using silica-based sol-gel stationary phases from methyl- and ethyl-trimethoxysilane precursors [J].
Giordano, Braden C. ;
Terray, Alex ;
Collins, Greg E. .
ELECTROPHORESIS, 2006, 27 (21) :4295-4302
[10]   Reversed-phase liquid chromatography on a microchip with sample injector and monolithic silica column [J].
Ishida, Akihiko ;
Yoshikawa, Takahiro ;
Natsume, Masamichi ;
Kamidate, Tamio .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1132 (1-2) :90-98