Quantification of lactate in synovia by microchip with contactless conductivity detection

被引:22
作者
Song, Zhirui [1 ]
Xu, Yue [2 ]
Chen, Zuanguang [1 ]
Yang, Jianping [1 ]
Li, Xinchun [1 ]
Zhang, Zhiguang [2 ]
机构
[1] Sun Yat Sen Univ, Sch Pharmaceut Sci, Guangzhou 510006, Guangdong, Peoples R China
[2] Sun Yat Sen Univ, Guanghua Sch Stomatol, Guangzhou 510055, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Microchip; Capillary electrophoresis; Contactless conductivity detection; Synovia; Lactate; CAPILLARY-ELECTROPHORESIS; LACTIC-ACID; AMPEROMETRIC BIOSENSOR; DIAGNOSIS; OXIDASE; PLASMA; SENSOR; GLASS; CHIP; HPLC;
D O I
10.1016/j.ab.2012.11.006
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This article describes the determination of lactate in synovia by microchip capillary electrophoresis (MCE) integrated with contactless conductivity detection (CCD). The optimal running buffer consists of 10 mM tris(hydroxymethyl)aminomethane, 1 mM HCl, and 0.1 mM hexadecyltrimethylammonium bromide (pH 9.1). The quantitative measurement of lactate in dilute synovia samples can be finished in less than 40 s. The results indicated that the peak area had a good linear relationship with lactate concentration in the range of 20 to 1000 mu M, and the correlation coefficient was 0.9984. The average recovery was from 96.6% to 106.1%, and the interday relative standard deviation was less than 4.0% (n = 6). The limit of detection (signal/noise = 3) reached 6.5 mu M. To validate the assay results, we compared the current method with the high-performance liquid chromatography method by measuring lactate in synovia samples. The data analysis verified that there was no significant difference between the two methods. Due to significant features such as low cost, integration, and miniaturization, the MCE CCD method may have great potential in clinical diagnosis. (C) 2012 Elsevier Inc. All rights reserved.
引用
收藏
页码:73 / 77
页数:5
相关论文
共 32 条
[11]   On-chip contactless four-electrode conductivity detection for capillary electrophoresis devices [J].
Laugere, F ;
Guijt, RM ;
Bastemeijer, J ;
van der Steen, G ;
Berthold, A ;
Baltussen, E ;
Sarro, P ;
van Dedem, GWK ;
Vellekoop, M ;
Bossche, A .
ANALYTICAL CHEMISTRY, 2003, 75 (02) :306-312
[12]   Cobalt hexacyanoferrate modified multi-walled carbon nanotubes/graphite composite electrode as electrochemical sensor on microfluidic chip [J].
Li, Xinchun ;
Chen, Zuanguang ;
Zhong, Yuwen ;
Yang, Fan ;
Pan, Jianbin ;
Liang, Yajing .
ANALYTICA CHIMICA ACTA, 2012, 710 :118-124
[13]   A Novel Immobilization Enzyme Lactate Fluorescence Capillary Biosensor [J].
Li Yong-Sheng ;
Ju Xiang ;
Gao Xiu-Feng ;
Yang Wei .
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2009, 37 (05) :637-642
[14]   Dual fluorescence/contactless conductivity detection for microfluidic chip [J].
Liu, Cui ;
Mo, Yun-yan ;
Chen, Zuan-guang ;
Li, Xiang ;
Li, Ou-lian ;
Zhou, Xie .
ANALYTICA CHIMICA ACTA, 2008, 621 (02) :171-177
[15]   LACTATE LEVELS IN BRUCELLA ARTHRITIS [J].
MAVRIDIS, AK ;
DROSOS, AA ;
TSOLAS, O ;
MOUTSOPOULOS, HM .
RHEUMATOLOGY INTERNATIONAL, 1984, 4 (04) :169-171
[16]   Determination of lactic acid in human plasma by HPLC with electrochemical detection [J].
Miyaguchi, Y ;
Kotani, A ;
Kusu, F .
BUNSEKI KAGAKU, 2002, 51 (09) :703-706
[17]   FLUOROMETRIC METHODS FOR THE MEASUREMENT OF INTERMEDIATE METABOLITES (LACTATE, PYRUVATE, ALANINE BETA-HYDROXYBUTYRATE, GLYCEROL) USING A COBAS FARA CENTRIFUGAL ANALYZER [J].
MONTI, LD ;
SANDOLI, PE ;
COSTA, S ;
PHAN, VC ;
PIATTI, PM .
JOURNAL OF AUTOMATIC CHEMISTRY, 1993, 15 (05) :177-181
[18]   Fabrication and characterization of disposable type lactate oxidase sensors for dairy products and clinical analysis [J].
Patel, NG ;
Erlenkötter, A ;
Cammann, K ;
Chemnitius, GC .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 67 (1-2) :134-141
[19]   Flow injection determination of lactate based on a photochemical reaction using photometric and chemiluminescence detection [J].
Pérez-Ruiz, T ;
Martínez-Lozano, C ;
Tomás, V ;
Martín, J .
ANALYST, 1999, 124 (10) :1517-1521
[20]   Capillary electrophoresis with contactless conductivity detection for the determination of carnitine and acylcarnitines in clinical samples [J].
Pormsila, Worapan ;
Morand, Rejane ;
Kraehenbuehl, Stephan ;
Hauser, Peter C. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (13-14) :921-926