Coordinating Retrieval and Register Studies Improves Postmarket Surveillance

被引:8
作者
Ellison, Peter [1 ]
Hallan, Geir [2 ]
Hol, Paul Johan [1 ]
Gjerdet, Nils Roar [3 ]
Havelin, Leif I. [2 ]
机构
[1] Univ Bergen, Dept Surg Sci, N-5021 Bergen, Hordaland, Norway
[2] Haukeland Hosp, Dept Orthopaed Surg, N-5021 Bergen, Norway
[3] Univ Bergen, Dept Clin Dent, N-5021 Bergen, Hordaland, Norway
关键词
NORWEGIAN ARTHROPLASTY REGISTER; TOTAL HIP REPLACEMENTS; FEMORAL COMPONENTS; EARLY REVISION; STEM DESIGNS; PROSTHESES; INTERFACE; CHARNLEY; CEMENT; IMPLANTS;
D O I
10.1007/s11999-012-2430-4
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background The relative risk of revision of the Titan (R) femoral stem due to aseptic loosening increased after 2000; however, the reasons for this have not been established. A retrieval analysis was initiated with the aim of delineating the failure mechanism. Questions/Purposes We asked whether aseptic loosening in stems after 2000 was associated with (1) appearance of osteolytic lesions, (2) wear particle exposure, (3) stem damage, or (4) changes to the implant or surgical instrumentation. Methods Femoral stems, cement, tissue, and radiographs were collected from 28 patients. We assessed the development of osteolytic lesions in 17 patients. Exposure to wear particles was quantified in 18 patients. Stem damage was assessed in 15 patients. We observed differences in the implants by examination of 24 retrieved stems. Information concerning changes to instrumentation was requested from the manufacturer. Results We found osteolysis in all patients receiving implants after 2000, which was associated with a median dose of cement and stem particles of 14,726/mm(2). Abrasion covered 59% of the surface of stems implanted from 1999. We identified geometric changes to the stem, the percent weight of aluminum in the stem's oxide layer decreased from 25% to 14% after 1997 and the rasp used to prepare the femoral cavity changed to a broach in 1999. Conclusions Stems implanted from 2000 failed through osteolysis induced by particles released from the cement and implant. Changes to implant geometry, surface oxide layer, and surgical tools occurred in the same time frame as the reduction in survivorship.
引用
收藏
页码:2995 / 3002
页数:8
相关论文
共 24 条
[1]
LOCALIZED ENDOSTEAL BONE LYSIS IN RELATION TO THE FEMORAL COMPONENTS OF CEMENTED TOTAL HIP ARTHROPLASTIES [J].
ANTHONY, PP ;
GIE, GA ;
HOWIE, CR ;
LING, RSM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (06) :971-979
[2]
PREPARATION AND SURFACE SPECTROSCOPIC CHARACTERIZATION OF OXIDE-FILMS ON TI6A14V [J].
ASK, M ;
LAUSMAA, J ;
KASEMO, B .
APPLIED SURFACE SCIENCE, 1989, 35 (03) :283-301
[3]
Baumann B, 2007, ORTHOPEDICS, V30, P551, DOI 10.3928/01477447-20070701-01
[4]
Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience [J].
Boyer, Patrick ;
Lazennec, Jean-Yves ;
Poupon, Joel ;
Rousseau, Marc-Antoine ;
Ravaud, Philippe ;
Catonne, Yves .
INTERNATIONAL ORTHOPAEDICS, 2009, 33 (05) :1209-1215
[5]
Early revision among 12,179 hip prostheses - A comparison of 10 different brands reported to the Norwegian Arthroplasty Register, 1987-1993 [J].
Espehaug, B ;
Havelin, LI ;
Engesaeter, LB ;
Vollset, SE ;
Langeland, N .
ACTA ORTHOPAEDICA SCANDINAVICA, 1995, 66 (06) :487-493
[6]
18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register Concerns about some newer implants [J].
Espehaug, Birgitte ;
Furnes, Ove ;
Engesaeter, Lars B. ;
Havelin, Leif I. .
ACTA ORTHOPAEDICA, 2009, 80 (04) :402-412
[7]
In vitro interface and cement mantle analysis of different femur stem designs [J].
Gravius, S. ;
Wirtz, D. C. ;
Siebert, C. H. ;
Andereya, St. ;
Mueller-Rath, R. ;
Maus, U. ;
Mumme, T. .
JOURNAL OF BIOMECHANICS, 2008, 41 (09) :2021-2028
[8]
Is there still a place for the cemented titanium femoral stem? 10,108 cases from the Norwegian Arthroplasty Register [J].
Hallan, Geir ;
Espehaug, Birgitte ;
Furnes, Ove ;
Wangen, Helge ;
Hol, Paul J. ;
Ellison, Peter ;
Havelin, Leif I. .
ACTA ORTHOPAEDICA, 2012, 83 (01) :1-6
[9]
EARLY ASEPTIC LOOSENING OF UNCEMENTED FEMORAL COMPONENTS IN PRIMARY TOTAL HIP-REPLACEMENT - A REVIEW BASED ON THE NORWEGIAN ARTHROPLASTY REGISTER [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (01) :11-17
[10]
THE EFFECT OF THE TYPE OF CEMENT ON EARLY REVISION OF CHARNLEY TOTAL HIP PROSTHESES - A REVIEW OF 8,579 PRIMARY ARTHROPLASTIES FROM THE NORWEGIAN ARTHROPLASTY REGISTER [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (10) :1543-1550