High-speed and high-resolution UPLC separation at zero degrees Celsius

被引:278
作者
Wales, Thomas E. [1 ]
Fadgen, Keith E. [3 ]
Gerhardt, Geoff C. [3 ]
Engen, John R. [1 ,2 ]
机构
[1] Northeastern Univ, Barnett Inst, Boston, MA 02115 USA
[2] Northeastern Univ, Dept Chem & Chem Biol, Boston, MA 02115 USA
[3] Waters Corp, Milford, MA 01757 USA
关键词
D O I
10.1021/ac8008862
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The conformational properties of proteins can be probed with hydrogen/deuterium exchange mass spectrometry (HXMS). In order to maintain the deuterium label during LC/MS analyses, chromatographic separation must be done rapidly (usually in under 8-10 min) and at 0 degrees C. Traditional RP-HPLC with similar to 3-mu m particles has shown generally poor chromatographic performance under these conditions and thereby has been prohibitive for HXMS analyses of larger proteins and many protein complexes. Ultraperformance liquid chromatography (UPLC) employs particles smaller than 2 mu m in diameter to achieve superior resolution, speed, and sensitivity as compared to HPLC. UPLC has previously been shown to be compatible with the fast separation and low temperature requirements of HXMS. Here we present construction and validation of a custom UPLC system for HXMS. The system is based on the Waters nanoACQUITY platform and contains a Peltier-cooled module that houses the injection and switching valves, online pepsin digestion column, and C-18 analytical separation column. Single proteins in excess of 95 kDa and a four-protein mixture in excess of 250 kDa have been used to validate the performance of this new system. Near-baseline resolution was achieved in 6-min separations at 0 degrees C and displayed a median chromatographic peak width of similar to 2.7 s at half-height. Deuterium recovery was similar to that obtained using a conventional HPLC and ice bath. This new system represents a significant advancement in HXMS technology that is expected to make the technique more accessible and mainstream in the near future.
引用
收藏
页码:6815 / 6820
页数:6
相关论文
共 10 条
[1]  
Chalmers Michael J, 2007, J Biomol Tech, V18, P194
[2]   Probing protein ligand interactions by automated hydrogen/deuterium exchange mass spectrometry [J].
Chalmers, MJ ;
Busby, SA ;
Pascal, BD ;
He, YJ ;
Hendrickson, CL ;
Marshall, AG ;
Griffin, PR .
ANALYTICAL CHEMISTRY, 2006, 78 (04) :1005-1014
[3]   Methods to study protein dynamics and folding by mass spectrometry [J].
Eyles, SJ ;
Kaltashov, IA .
METHODS, 2004, 34 (01) :88-99
[4]   Protein analysis by hydrogen exchange mass spectrometry [J].
Hoofnagle, AN ;
Resing, KA ;
Ahn, NG .
ANNUAL REVIEW OF BIOPHYSICS AND BIOMOLECULAR STRUCTURE, 2003, 32 :1-25
[5]   Ultra-performance liquid chromatography coupled to quadrupole-orthogonal time-of-flight mass spectrometry [J].
Plumb, R ;
Castro-Perez, J ;
Granger, J ;
Beattie, I ;
Joncour, K ;
Wright, A .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (19) :2331-2337
[6]   Hydrogen exchange mass spectrometry for the analysis of protein dynamics [J].
Wales, TE ;
Engen, JR .
MASS SPECTROMETRY REVIEWS, 2006, 25 (01) :158-170
[7]   Hydrogen exchange-mass spectrometry - Optimization of digestion conditions [J].
Wang, LT ;
Pan, H ;
Smith, DL .
MOLECULAR & CELLULAR PROTEOMICS, 2002, 1 (02) :132-138
[8]   Extensive deuterium back-exchange in certain immobilized pepsin columns used for H/D exchange mass spectrometry [J].
Wu, Y ;
Kaveti, S ;
Engen, JR .
ANALYTICAL CHEMISTRY, 2006, 78 (05) :1719-1723
[9]   Ultra performance liquid chromatography (UPLC) further improves hydrogen/deuterium exchange mass spectrometry [J].
Wu, Y ;
Engen, JR ;
Hobbins, WB .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (02) :163-167
[10]  
ZHANG ZQ, 1993, PROTEIN SCI, V2, P522