Theoretical Comparison of the Iodine Quantification Accuracy of Two Spectral CT Technologies

被引:16
作者
Bornefalk, Hans [1 ]
Persson, Mats [1 ]
机构
[1] AlbaNova Univ Ctr, Royal Inst Technol, Dept Phys, SE-10691 Stockholm, Sweden
关键词
Dual energy computed tomography (CT); iodine quantification; photon counting multibin CT; spectral CT; DUAL-ENERGY CT; COMPUTED-TOMOGRAPHY; MATERIAL DECOMPOSITION; NOISE; PERFORMANCE; STATISTICS; SYSTEM; SNR;
D O I
10.1109/TMI.2013.2290198
中图分类号
TP39 [计算机的应用];
学科分类号
080201 [机械制造及其自动化];
摘要
We compare the theoretical limits of iodine quantification for the photon counting multibin and dual energy technologies. Dual energy systems by necessity have to make prior assumptions in order to quantify iodine. We explicitly allow the multibin system to make the same assumptions and also allow them to be wrong. We isolate the effect of technology from imperfections and implementation issues by assuming both technologies to be ideal, i.e., without scattered radiation, unity detection efficiency and perfect energy response functions, and by applying the Cramer-Rao lower bound methodology to assess the quantification accuracy. When priors are wrong the maximum likelihood estimates will be biased and the mean square error of the quantification error is a more appropriate figure of merit. The evaluation assumes identical X-ray spectra for both methodologies and for that reason a sensitivity analysis is performed with regard to the assumed X-ray spectrum. We show that when iodine is quantified over regions of interest larger than 6 cm, multibin systems benefit by independent estimation of three basis functions. For smaller regions of interest multibin systems can increase quantification accuracy by making the same prior assumptions as dual energy systems.
引用
收藏
页码:556 / 565
页数:10
相关论文
共 44 条
[21]
Relationship between patient centering, mean computed tomography numbers and noise in abdominal computed tomography: Influence of anthropomorphic parameters [J].
Kim, Mi Sung ;
Singh, Sarabjeet ;
Halpern, Elkan ;
Saini, Sanjay ;
Kalra, Mannudeep K. .
WORLD JOURNAL OF RADIOLOGY, 2012, 4 (03) :102-108
[22]
Krauss B, 2011, MED RADIOL DIAGN IMA, P11, DOI 10.1007/174_2010_44
[23]
Least squares parameter estimation methods for material decomposition with energy discriminating detectors [J].
Le, Huy Q. ;
Molloi, Sabee .
MEDICAL PHYSICS, 2011, 38 (01) :245-255
[24]
GENERALIZED IMAGE COMBINATIONS IN DUAL KVP DIGITAL RADIOGRAPHY [J].
LEHMANN, LA ;
ALVAREZ, RE ;
MACOVSKI, A ;
BRODY, WR ;
PELC, NJ ;
RIEDERER, SJ ;
HALL, AL .
MEDICAL PHYSICS, 1981, 8 (05) :659-667
[25]
Lehmann LA, 1986, DIGITAL RADIOGRAPHY, P145
[26]
Quantitative imaging of element composition and mass fraction using dual-energy CT: Three-material decomposition [J].
Liu, Xin ;
Yu, Lifeng ;
Primak, Andrew N. ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2009, 36 (05) :1602-1609
[27]
The potential of dual-energy computed tomography for quantitative decomposition of soft tissues to water, protein and lipid in brachytherapy [J].
Malusek, A. ;
Karlsson, M. ;
Magnusson, M. ;
Carlsson, G. Alm .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (04) :771-785
[28]
Dual-Source Dual-Energy CT With Additional Tin Filtration: Dose and Image Quality Evaluation in Phantoms and In Vivo [J].
Primak, Andrew N. ;
Giraldo, Juan Carlos Ramirez ;
Eusemann, Christian D. ;
Schmidt, Bernhard ;
Kantor, Birgit ;
Fletcher, Joel G. ;
McCollough, Cynthia H. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (05) :1164-1174
[29]
NOISE POWER SPECTRUM IN COMPUTED X-RAY TOMOGRAPHY [J].
RIEDERER, SJ ;
PELC, NJ ;
CHESLER, DA .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (03) :446-454
[30]
K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors [J].
Roessl, E. ;
Proksa, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15) :4679-4696