Relationship between patient centering, mean computed tomography numbers and noise in abdominal computed tomography: Influence of anthropomorphic parameters

被引:10
作者
Kim, Mi Sung [1 ]
Singh, Sarabjeet [2 ]
Halpern, Elkan [2 ]
Saini, Sanjay [2 ]
Kalra, Mannudeep K. [2 ]
机构
[1] Kwandong Univ, Myongji Hosp, 522 Naegok Dong, Gangneung Si 220801, Gangwon Do, South Korea
[2] Harvard Med Sch, Massachusetts Gen Hosp, Dept Radiol, 55 Fruit St, Boston, MA 02114 USA
关键词
Automatic exposure control; Computed tomography radiation dose reduction; Patient offcentering; Tube current modulation;
D O I
10.4329/wjr.v4.i3.102
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
AIM: To determine the influence of anthropomorphic parameters on the relationship between patient centering, mean computed tomography (CT) numbers and quantitative image noise in abdominal CT. METHODS: Our Institutional Review Board approved study included 395 patients (age range 21-108, years; male: female = 195:200) who underwent contrast-enhanced abdominal CT on a 16-section multi-detector row scanner (GE LightSpeed 16). Patient centering in the gantry isocenter was measured from the lateral localizer radiograph (off center S = patient off centered superior to isocenter; off center I = patient off centered inferior to isocenter). Mean CT numbers (Hounsfield Units: HU) and noise (standard deviation of CT numbers: SD) were measured in the anterior (aHU, aSD) and posterior (pHU, pSD) abdominal wall subcutaneous fat and liver parenchyma (LivHU, LivSD) at the level of the porta hepatis. Patients' age, gender, weight, body mass index and maximal anteroposterior diameter were recorded. The data were analyzed using linear regression analysis. RESULTS: Most patients (81%; 320/395) were not correctly centered in the gantry isocenter for abdominal CT scanning. Mean CT numbers in the abdominal wall increased significantly with an increase in the off-centering distance, regardless of the direction of the off-center (P < 0.05). There was a substantial increase in pSD (P = 0.01) and LivSD (P = 0.017) with off-centering. Change in mean CT numbers and image noise along the off-center distance was influenced by the patient size (P < 0.01). CONCLUSION: Inappropriate patient centering for CT scanning adversely affects the reliability of mean CT numbers and image noise. (c) 2012 Baishideng. All rights reserved.
引用
收藏
页码:102 / 108
页数:7
相关论文
共 11 条
[1]
Minimizing radiation dose for pediatric body applications of single-detector helical CT: Strategies at a large children's hospital [J].
Donnelly, LF ;
Emery, KH ;
Brody, AS ;
Laor, T ;
Gylys-Morin, VM ;
Anton, CG ;
Thomas, SR ;
Frush, DP .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :303-306
[2]
Radiation dose and cancer risk from pediatric CT examinations on 64-slice CT: A phantom study [J].
Feng, Shi-Ting ;
Law, Martin Wai-Ming ;
Huang, Bingsheng ;
Ng, Sherry ;
Li, Zi-Ping ;
Meng, Quan-Fei ;
Khong, Pek-Lan .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 76 (02) :E19-E23
[3]
Cumulative Radiation Exposure and Cancer Risk Estimates in Emergency Department Patients Undergoing Repeat or Multiple CT [J].
Griffey, Richard T. ;
Sodickson, Aaron .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) :887-892
[4]
Sixteen-detector row CT of abdomen and pelvis: Study for optimization of Z-axis modulation technique performed in 153 patients [J].
Kalra, MK ;
Maher, MM ;
Kamath, RS ;
Horiuchi, T ;
Toth, TL ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 233 (01) :241-249
[5]
Strategies for CT radiation dose optimization [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Hamberg, LM ;
Blake, MA ;
Shephard, JA ;
Saini, S .
RADIOLOGY, 2004, 230 (03) :619-628
[6]
Application of automatic vertical positioning software to reduce radiation exposure in multidetector row computed tomography of the chest [J].
Li, Jianhai ;
Udayasankar, Unni K. ;
Toth, Thomas L. ;
Small, William C. ;
Kalra, Mannudeep K. .
INVESTIGATIVE RADIOLOGY, 2008, 43 (06) :447-452
[7]
Automatic patient centering for MDCT: Effect on radiation dose [J].
Li, Jianhai ;
Udayasankar, Unni K. ;
Toth, Thomas L. ;
Seamans, John ;
Small, William C. ;
Kalra, Mannudeep K. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (02) :547-552
[8]
Analysis of surface dose variation in CT procedures [J].
Lucas, PA ;
Castellano, IA ;
Dance, DR ;
Carruana, EV .
BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (888) :1128-1136
[9]
Application of rational practice and technical advances for optimizing radiation dose for chest CT [J].
Maher, MM ;
Kalra, MK ;
Toth, TL ;
Wittram, C ;
Saini, S ;
Shepard, J .
JOURNAL OF THORACIC IMAGING, 2004, 19 (01) :16-23
[10]
Misoperation of CT Automatic Tube Current Modulation Systems with Inappropriate Patient Centering: Phantom Studies [J].
Matsubara, Kosuke ;
Koshida, Kichiro ;
Ichikawa, Katsuhiro ;
Suzuki, Masayuki ;
Takata, Tadanori ;
Yamamoto, Tomoyuki ;
Matsui, Osamu .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) :862-865