Application of automatic vertical positioning software to reduce radiation exposure in multidetector row computed tomography of the chest

被引:19
作者
Li, Jianhai [1 ]
Udayasankar, Unni K. [1 ]
Toth, Thomas L. [2 ]
Small, William C. [1 ]
Kalra, Mannudeep K. [3 ]
机构
[1] Emory Univ, Sch Med, Dept Radiol, Atlanta, GA 30322 USA
[2] Gen Elect Healthcare Technol, Waukesha, WI USA
[3] Massachusetts Gen Hosp, Dept Radiol, Boston, MA 02114 USA
关键词
multidetector row computed tomography; radiation; automatic vertical positioning; patient mispositioning;
D O I
10.1097/RLI.0b013e318169005e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To determine the effect of automatic vertical positioning (AVP) software on radiation dose reduction in multidetector row computed tomography of the chest. Materials and Methods: Two hundred forty-three consecutive patients (mean age, 54; range, 20-93 years), who underwent 64-slice multidetector row computed tomography of the chest from November 2005 to April 2006, were included in this study. This study was approved by our institutional review board with waiver of informed consent. Technologists initially positioned the patient in the gantry by adjusting the table height (vertical positioning) using a laser guidance system and patient surface anatomy. AVP software was then used to determine the true vertical positioning of the patient based on matching the individual patient's mean centroid (center of mass) throughout the chest as determined by the lateral projection with the gantry isocenter and center of the bow-tie filter. The resultant reduction in surface radiation dose with this attenuation-based vertical positioning was then calculated. Software-determined changes in patient vertical positioning and resultant radiation dose savings were analyzed using the t test and Pearson correlation statistics. Results: Vertical positioning of the patients by the technologist differed from AVP with the bow-tie filter in 95.5% (232 of 243) of patients with an average vertical mispositioning distance of 33.2 +/- 1.1 mm (range, 5.1-97 mm). Among mispositioned patients, 97% (225 of 232) were vertically mispositioned below the isocenter with a mean distance of 33.8 +/- 1.1 turn (range, 5.1-97 mm) and 3% (7 of 232) patients were vertically mispositioned above the isocenter, with a mean distance of 12.7 +/- 2.8 mm (range, 6-28 mm). The average surface radiation dose reduction with AVP software was 19% +/- 1% (range, 1%-46.6%). There was a strong correlation between vertical mispositioning distance and the surface radiation dose reduction facilitated by AVP software (r(2) = 0.9, P < 0.001). Conclusion: AVP software can help in optimum patient vertical positioning leading to substantial reduction in surface radiation dose.
引用
收藏
页码:447 / 452
页数:6
相关论文
共 21 条
[1]   Automatic tube-current modulation in CT -: A comparison between different solutions [J].
Althén, JN .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :308-312
[2]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[3]   Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography [J].
Einstein, Andrew J. ;
Henzlova, Milena J. ;
Rajagopalan, Sanjay .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2007, 298 (03) :317-323
[4]   Evaluation of a reduced dose protocol for respiratory gated lung computed tomography in an animal model [J].
Elgeti, Thomas ;
Proquitte, Hans ;
Rogalla, Noga E. ;
Mews, Juergen ;
Hamm, Bernd ;
Schmalisch, Gerd ;
Rogalla, Patrik .
INVESTIGATIVE RADIOLOGY, 2007, 42 (04) :230-234
[5]   ESTIMATION OF RISK OF CANCER INDUCTION IN POPULATIONS EXPOSED TO LOW-LEVEL RADIATION [J].
FABRIKANT, JI .
INVESTIGATIVE RADIOLOGY, 1982, 17 (04) :342-349
[6]  
HUNCT W, 2005, J COMPUT ASSIST TOMO, V29, P140
[7]   Chest CT performed with z-axis modulation: Scanning protocol and radiation dose [J].
Kalra, MK ;
Rizzo, S ;
Maher, MM ;
Halpern, EF ;
Toth, TL ;
Shepard, JAO ;
Aquino, SL .
RADIOLOGY, 2005, 237 (01) :303-308
[8]   Sixteen-detector row CT of abdomen and pelvis: Study for optimization of Z-axis modulation technique performed in 153 patients [J].
Kalra, MK ;
Maher, MM ;
Kamath, RS ;
Horiuchi, T ;
Toth, TL ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 233 (01) :241-249
[9]   Automatic patient centering for MDCT: Effect on radiation dose [J].
Li, Jianhai ;
Udayasankar, Unni K. ;
Toth, Thomas L. ;
Seamans, John ;
Small, William C. ;
Kalra, Mannudeep K. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (02) :547-552
[10]   Radiation exposure at chest CT:: A statement of the Fleischner Society [J].
Mayo, JR ;
Aldrich, J ;
Müller, NL .
RADIOLOGY, 2003, 228 (01) :15-21