Automatic MR volume registration and its evaluation for the pelvis and prostate

被引:70
作者
Fei, BW [1 ]
Wheaton, A
Lee, ZH
Duerk, JL
Wilson, DL
机构
[1] Case Western Reserve Univ, Dept Biomed Engn, Cleveland, OH 44106 USA
[2] Univ Hosp Cleveland, Dept Radiol, Cleveland, OH 44106 USA
[3] Case Western Reserve Univ, Cleveland, OH 44106 USA
关键词
D O I
10.1088/0031-9155/47/5/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A three-dimensional (3D) mutual information registration method was created and used to register MRI volumes of the pelvis and prostate. It had special features to improve robustness. First, it used a multi-resolution approach and performed registration from low to high resolution. Second, it used two similarity measures, correlation coefficient at lower resolutions and mutual information at full resolution, because of their particular advantages. Third, we created a method to avoid local minima by restarting the registration with randomly perturbed parameters. The criterion for restarting was a correlation coefficient below an empirically determined threshold. Experiments determined the accuracy of registration under conditions found in potential applications in prostate cancer diagnosis, staging, treatment and interventional MRI (iMRI) guided therapies. Images were acquired in the diagnostic (supine) and treatment position (supine with legs raised). Images were also acquired as a function of bladder filling and the time interval between imaging sessions. Overall studies on three patients and three healthy volunteers, when both volumes in a pair were obtained in the diagnostic position under comparable conditions, bony landmarks and prostate 3D centroids were aligned within 1.6 +/- 0.2 mm and 1.4 +/- 0.2 mm, respectively, values only slightly larger than a voxel. Analysis suggests that actual errors are smaller because of the uncertainty in landmark localization and prostate segmentation. Between the diagnostic and treatment positions, bony landmarks continued to register well, but prostate centroids moved towards the posterior 2.8-3.4 mm. Manual cropping to remove voxels in the legs was necessary to register these images. In conclusion, automatic, rigid body registration is probably sufficiently accurate 0 for Man. applications in prostate cancer. For potential iMRI-guided treatments, the small prostate displacement between the diagnostic and treatment positions can probably be avoided by acquiring volumes in similar positions and by reducing bladder and rectal volumes.
引用
收藏
页码:823 / 838
页数:16
相关论文
共 39 条
  • [1] Prostate target volume variations during a course of radiotherapy
    Antolak, JA
    Rosen, II
    Childress, CH
    Zagars, GK
    Pollack, A
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03): : 661 - 672
  • [2] MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS
    BALTER, JM
    SANDLER, HM
    LAM, K
    BREE, RL
    LICHTER, AS
    TENHAKEN, RK
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01): : 113 - 118
  • [3] OPTIMIZATION OF PROSTATE CARCINOMA STAGING - COMPARISON OF IMAGING AND CLINICAL METHODS
    BONI, RAH
    BONER, JA
    DEBATIN, JF
    TRINKLER, F
    KNONAGEL, H
    VONHOCHSTETTER, A
    HELFENSTEIN, U
    KRESTIN, GP
    [J]. CLINICAL RADIOLOGY, 1995, 50 (09) : 593 - 600
  • [4] [MoO2(acac)2], a versatile precursor for diazenido- and hydrazido-complexes
    Carrillo, D
    [J]. COMPTES RENDUS DE L ACADEMIE DES SCIENCES SERIE II FASCICULE C-CHIMIE, 2000, 3 (03): : 175 - 181
  • [5] Robust registration method for interventional MRI-guided thermal ablation of prostate cancer
    Fei, BW
    Wheaton, A
    Lee, Z
    Nagano, K
    Duerk, JL
    Wilson, DL
    [J]. MEDICAL IMAGING 2001: VISUALIZATION, DISPLAY, AND IMAGE-GUIDED PROCEDURES, 2001, 4319 : 53 - 60
  • [6] Visual assessment of the accuracy of retrospective registration of MR and CT images of the brain
    Fitzpatrick, JM
    Hill, DLG
    Shyr, Y
    West, J
    Studholme, C
    Maurer, CR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (04) : 571 - 585
  • [7] Accurate registration of serial 3D MR brain images and its application to visualizing change in neurodegenerative disorders
    Freeborough, PA
    Woods, RP
    Fox, NC
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (06) : 1012 - 1022
  • [8] GRAY H, 1977, ANATOMY DESCRIPTIVE, P1010
  • [9] Hamilton RJ, 1999, J NUCL MED, V40, P347
  • [10] Herk M, 1995, INT J RADIAT ONCOL, V33, P1311