摘要
Corrosion has long been recognized to occur in total hip arthroplasty, but the local effects of this process have only recently become better understood. This article provides an overview of corrosion at modular junctions, and discusses the various etiologic factors for corrosion and the biologic response to metal debris released from this junction. Algorithms are provided for diagnosis and treatment, in accordance with the best available data.
引用
收藏
页码:9 / +
页数:11
相关论文
共 90 条
[1]
[Anonymous], J ARTHROPLASTY
[2]
A Comparison of Two Implant Systems in Restoration of Hip Geometry in Arthroplasty [J].
Archibeck, Michael J. ;
Cummins, Tamara ;
Carothers, Joshua ;
Junick, Daniel W. ;
White, Richard E., Jr. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (02) :443-446
[3]
Corrosion-Induced Fracture of a Double-Modular Hip Prosthesis A Case Report [J].
Atwood, Sara A. ;
Patten, Eli W. ;
Bozic, Kevin J. ;
Pruitt, Lisa A. ;
Ries, Michael D. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (06) :1522-1525
[4]
BOBYN JD, 1994, CLIN ORTHOP RELAT R, P27
[5]
High failure rates with a large-diameter hybrid metal-on-metal total hip replacement CLINICAL, RADIOLOGICAL AND RETRIEVAL ANALYSIS [J].
Bolland, B. J. R. F. ;
Culliford, D. J. ;
Langton, D. J. ;
Millington, J. P. S. ;
Arden, N. K. ;
Latham, J. M. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (05) :608-615
[6]
Trunion fracture of a fully porous-coated femoral stem - Case report [J].
Botti, TP ;
Gent, J ;
Martell, JM ;
Manning, DW .
JOURNAL OF ARTHROPLASTY, 2005, 20 (07) :943-945
[7]
FRETTING CORROSION ACCELERATES CREVICE CORROSION OF MODULAR HIP TAPERS [J].
BROWN, SA ;
FLEMMING, CAC ;
KAWALEC, JS ;
PLACKO, HE ;
VASSAUX, C ;
MERRITT, K ;
PAYER, JH ;
KRAAY, MJ .
JOURNAL OF APPLIED BIOMATERIALS, 1995, 6 (01) :19-26
[8]
In vitro comparison of frictional torque and torsional resistance of aged conventional gamma-in-nitrogen sterilized polyethylene versus aged highly crosslinked polyethylene articulating against head sizes larger than 32 mm [J].
Burroughs, Brian R. ;
Muratoglu, Orhun K. ;
Bragdon, Charles R. ;
Wannomae, Keith K. ;
Christensen, Steven ;
Lozynsky, Andrew J. ;
Harris, William H. .
ACTA ORTHOPAEDICA, 2006, 77 (05) :710-718
[9]
Femoral Stem Fracture and In Vivo Corrosion of Retrieved Modular Femoral Hips [J].
Carlson, J. Caitlin Huot ;
Van Citters, Douglas W. ;
Currier, John H. ;
Bryant, Amber M. ;
Mayor, Michael B. ;
Collier, John P. .
JOURNAL OF ARTHROPLASTY, 2012, 27 (07) :1389-1396
[10]
Chana R, 2012, J BONE JOINT SURG BR, V94B, P281, DOI [10.1302/0301-620X.94B2, 10.1302/0301-620X.94B2.27247]