Singles transmission scans performed post-injection for quantitative whole body PET imaging
被引:43
作者:
Smith, RJ
[1
]
Karp, JS
[1
]
Muehllehner, G
[1
]
Gualtieri, E
[1
]
Benard, F
[1
]
机构:
[1] UGM LAB,PHILADELPHIA,PA 19104
基金:
美国国家卫生研究院;
关键词:
D O I:
10.1109/23.597009
中图分类号:
TM [电工技术];
TN [电子技术、通信技术];
学科分类号:
0808 ;
0809 ;
摘要:
Post-injection singles transmission scanning has been implemented in the septumless PENN PET 240H scanner. The method uses a 6 mCi point transmission source of Cs-137 at the axial center and 36 cm off transaxial center of the camera field of view. Singles transmission scans of 1.8 minutes per bed axial position provide similar scan count densities to 15 minute coincidence transmission scans with a 0.5 mCi Ge-68 rod transmission source. Scatter and emission contamination suppression are achieved by applying a narrow 662 keV transmission photopeak energy window. The residual 511 keV emission contamination constitutes a background of uniform spatial distribution. Accurate and low noise attenuation correction is achieved by segmenting the singles transmission images into lung and soft tissue volumes. Established 511 keV gamma ray attenuation coefficients are then applied and these images are forward projected for attenuation correction. Expectation maximisation with ordered subsets (OS-EM) reconstruction is performed for the emission data to improve image quality. Both the segmentation and OS-EM reconstruction maintain quantitative accuracy in the fully corrected emission images compared to measured coincidence transmission correction. Thus, a clinical protocol involving 40 minutes of emission scans followed by 20 minutes of singles transmission scans allow the 60 cm of the human torso to be fully scanned within 60 minutes. These quantitative whole body FDG PET images may then be used for tumor grading and assessment of tumor response to treatment.
引用
收藏
页码:1329 / 1335
页数:7
相关论文