Film dosimetry for intensity modulated radiation therapy: Dosimetric evaluation

被引:76
作者
Ju, SG [1 ]
Ahn, YC [1 ]
Huh, SJ [1 ]
Yeo, IJ [1 ]
机构
[1] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Radiat Oncol, Seoul 135710, South Korea
关键词
film dosimetry; penumbra effect; lead filter; IMRT;
D O I
10.1118/1.1449493
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray film has been used for the dosimetry of intensity modulated radiation therapy (IMRT). However, the over-response of the film to low-energy photons is a significant problem in photon beam dosimetry, especially in regions outside penumbra. In IMRT, the radiation field consists of multiple small fields and their outside-penumbra regions; thus, the film dosimetry, for it involves the source of over-response in its radiation field. In this study we aim to verify and possibly improve film dosimetry for IMRT. Two types of modulated beams were constructed by combining five to seven different static radiation fields using 6 MV x rays. For verifying film dosimetry, x-ray films and an ion chamber were used to measure dose profiles at various depths in a phantom. The film setups include both parallel and perpendicular arrangements against the beam incident direction. In addition. to reduce an over-response, we placed 0.01 in. (0.25 mm) thick lead filters on both sides of the film. Compared with ion-chamber measurement, measured dose profiles showed the film over-response at outside-penumbra and low-dose regions. The error increased with depths and approached 15% as a maximum for the field size of 15 cm X 15 cm at 10 cm depth. The use of filters reduced the error down to 3%. In this study we demonstrated that film dosimetry for IMRT involves sources of error due to its over-response to low-energy photons, with the error most transparent in the low-dose region. The use of filters could enhance the accuracy in film dosimetry for IMRT. In this regard, the use of an optimal filter condition is recommended. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:351 / 355
页数:5
相关论文
共 7 条
[1]   A new approach to film dosimetry for high energy photon beams: Lateral scatter filtering [J].
Burch, SE ;
Kearfott, KJ ;
Trueblood, JH ;
Sheils, WC ;
Yeo, JI ;
Wang, CKC .
MEDICAL PHYSICS, 1997, 24 (05) :775-783
[2]   Planning, delivery, and quality assurance of intensity-modulated radiotherapy using dynamic multileaf collimator: A strategy for large-scale implementation for the treatment of carcinoma of the prostate [J].
Burman, C ;
Chui, CS ;
Kutcher, G ;
Leibel, S ;
Zelefsky, M ;
LoSasso, T ;
Spirou, S ;
Wu, QW ;
Yang, J ;
Stein, J ;
Mohan, R ;
Fuks, Z ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (04) :863-873
[3]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[4]   Quantitative dosimetric verification of an IMRT planning and delivery system [J].
Low, DA ;
Mutic, S ;
Dempsey, JF ;
Gerber, RL ;
Bosch, WR ;
Perez, CA ;
Purdy, JA .
RADIOTHERAPY AND ONCOLOGY, 1998, 49 (03) :305-316
[5]   Initial experience with intensity-modulated conformal radiation therapy for treatment of the head and neck region [J].
Verellen, D ;
Linthout, N ;
VandenBerge, D ;
Bel, A ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (01) :99-114
[6]   Dosimetric verification of intensity-modulated fields [J].
Wang, XH ;
Spirou, S ;
LoSasso, T ;
Stein, J ;
Chui, CS ;
Mohan, R .
MEDICAL PHYSICS, 1996, 23 (03) :317-327
[7]   A filtration method for improving film dosimetry in photon radiation therapy [J].
Yeo, IJ ;
Wang, CKC ;
Burch, SE .
MEDICAL PHYSICS, 1997, 24 (12) :1943-1953