Addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain

被引:66
作者
Britten, AJ [1 ]
Crotty, M
Kiremidjian, H
Grundy, A
Adam, EJ
机构
[1] St Georges Healthcare NHS Trust, Dept Phys Med, London SW17 0QT, England
[2] St Georges Healthcare NHS Trust, Dept Diagnost Radiol, London SW17 0QT, England
关键词
D O I
10.1259/bjr/78576048
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study validates a method to add spatially correlated statistical noise to an image, applied to transaxial X-ray CT images of the head to simulate exposure reduction by up to 50%. 23 patients undergoing routine head CT had three additional slices acquired for validation purposes, two at the same clinical 420 mAs exposure and one at 300 mAs. Images at the level of the cerebrospinal fluid filled ventricles gave readings of noise from a single image, with subtraction of image pairs to obtain noise readings from non-uniform tissue regions. The spatial correlation of the noise was determined and added to the acquired 420 mAs image to simulate images at 340 mAs, 300 mAs, 260 mAs and 210 mAs. Two radiologists assessed the images, finding little difference between the 300 mAs simulated and acquired images. The presence of periventricular low density lesions (PVLD) was used as an example of the effect of simulated dose reduction oil diagnostic accuracy. and Visualization of the internal capsule was Used as a measure of image quality. Diagnostic accuracy for the diagnosis of PVLD did not fall significantly even down to 210 mAs, though Visualization of the internal capsule was poorer at lower exposure. Further work is needed to investigate means of measuring statistical noise without the need for uniform tissue areas, or image pairs. This technique has been shown to allow sufficiently accurate simulation Of dose reduction and image quality degradation, even when the statistical noise is spatially correlated.
引用
收藏
页码:323 / 328
页数:6
相关论文
共 12 条
[1]  
BOGARTZ G, 16262 EUR EN
[2]  
EVANS AL, 1981, EVALUATION MED IMAGE, P32
[3]   Computer-simulated radiation dose reduction for abdominal multidetector CT of pediatric patients [J].
Frush, DP ;
Slack, CC ;
Hollingsworth, CL ;
Bisset, GS ;
Donnelly, LF ;
Hsieh, J ;
Lavin-Wensell, T ;
Mayo, JR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (05) :1107-1113
[4]   THE EFFECT OF MAS VARIATION UPON COMPUTED-TOMOGRAPHY IMAGE QUALITY AS EVALUATED BY INVIVO AND INVITRO STUDIES [J].
HAAGA, JR ;
MIRALDI, F ;
MACINTYRE, W ;
LIPUMA, JP ;
BRYAN, PJ ;
WIESEN, E .
RADIOLOGY, 1981, 138 (02) :449-454
[5]  
HUGHES JS, 1993, RADIAT EXPOSURE UK P
[6]   CHRONIC INFILTRATIVE LUNG-DISEASE - COMPARISON OF DIAGNOSTIC ACCURACIES OF RADIOGRAPHY AND LOW-AND CONVENTIONAL-DOSE THIN-SECTION CT [J].
LEE, KS ;
PRIMACK, SL ;
STAPLES, CA ;
MAYO, JR ;
ALDRICH, JE ;
MULLER, NL .
RADIOLOGY, 1994, 191 (03) :669-673
[7]   Simulated dose reduction in conventional chest CT: Validation study [J].
Mayo, JR ;
Whittall, KP ;
Leung, AN ;
Hartman, TE ;
Park, CS ;
Primack, SL ;
Chambers, GK ;
Limkeman, MK ;
Toth, TL ;
Fox, SH .
RADIOLOGY, 1997, 202 (02) :453-457
[8]   GT OF THE CHEST - MINIMAL TUBE CURRENT REQUIRED FOR GOOD IMAGE QUALITY WITH THE LEAST RADIATION-DOSE [J].
MAYO, JR ;
HARTMAN, TE ;
LEE, KS ;
PRIMACK, SL ;
VEDAL, S ;
MULLER, NL .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (03) :603-607
[9]   LOW-DOSE CT OF THE LUNGS - PRELIMINARY-OBSERVATIONS [J].
NAIDICH, DP ;
MARSHALL, CH ;
GRIBBIN, C ;
ARAMS, RS ;
MCCAULEY, DI .
RADIOLOGY, 1990, 175 (03) :729-731
[10]   Radiation exposure and image quality in chest CT examinations [J].
Ravenel, J ;
Scalzetti, EM ;
Huda, W ;
Garrisi, W .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (02) :279-284