Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol

被引:168
作者
Coleman, A [1 ]
Freeman, P
Steel, S
Shennan, A
机构
[1] Guys & St Thomas NHS Fdn Trust, Rayne Inst, Dept Med Phys, London SE1 7EH, England
[2] Kings Coll London, Sch Med, London WC2R 2LS, England
关键词
automated devices; blood pressure monitoring; oscillometric measurement; validation studies;
D O I
10.1097/01.mbp.0000189788.05736.5f
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background The Omron 705IT (HEM-759-E, Omron Corporation, Kyoto, Japan) is an automated oscillometric upper arm blood pressure monitor for the professional and home use markets. The aim of this study was to validate the accuracy of this device according to the British Hypertension Society and the Association for the Advancement of Medical Instrumentation SP10 validation criteria. Methods Study participants were recruited until a total of 85 were obtained that filled the blood pressure categories specified by the British Hypertension Society protocol. Recruitment to the study was from the general medical and specialist clinics and from the staff at Guy's & St Thomas' Hospital in London, UK. Nine sequential same-arm blood pressure readings were taken from each participant by two trained observers, alternating between mercury reference sphygmomanometers and the Omron 705IT (HEM-759-E). The differences between the reference and test device readings, for both systolic and diastolic pressures, were compared with British Hypertension Society and Association for the Advancement of Medical Instrumentation criteria to determine the outcome of the study. Results The Omron 705IT (HEM-759-E) is graded 'A' for systolic and 'A' for diastolic blood pressures according to the British Hypertension Society criteria. The mean (standard deviation) of the difference between the observer and the device measurements was 0.60 (6-0) mmHg for systolic and -3.15 (6.6) mmHg for diastolic pressures, respectively. The device, therefore, also satisfies the Association for the Advancement of Medical Instrumentation SP10 standard, that requires differences of less than +/- 5 (8) mmHg. Conclusions The Omron 705IT (HEM-759-E) achieved an 'A/A' performance classification under the British Hypertension Society criteria and passes the Association for the Advancement of Medical Instrumentation requirements for the study population. It can be recommended for professional and home-use in an adult population.
引用
收藏
页码:27 / 32
页数:6
相关论文
共 11 条
[1]  
*ASS ADV MED INSTR, 2003, SP102002 ANSI AAMI
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]  
*BS, 1996, EN106011996 BS 1
[4]   Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care [J].
Coleman, AJ ;
Steel, SD ;
Ashworth, M ;
Vowler, SL ;
Shennan, A .
BLOOD PRESSURE MONITORING, 2005, 10 (04) :181-188
[5]   Validation of the Microlife BP3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol [J].
Cuckson, AC ;
Reinders, A ;
Shabeeh, H ;
Shennan, AH .
BLOOD PRESSURE MONITORING, 2002, 7 (06) :319-324
[6]   Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT [J].
El Assaad, MA ;
Topouchian, JA ;
Asmar, RG .
BLOOD PRESSURE MONITORING, 2003, 8 (03) :127-133
[7]   A validation of the Mobil O Graph (version 12) ambulatory blood pressure monitor [J].
Jones, CR ;
Taylor, K ;
Chowienczyk, P ;
Poston, L ;
Shennan, AH .
BLOOD PRESSURE MONITORING, 2000, 5 (04) :233-238
[8]   A website for blood pressure measuring devices: dableducational.com [J].
O'Brien, E .
BLOOD PRESSURE MONITORING, 2003, 8 (04) :177-180
[9]   Replacing the mercury sphygmomanometer [J].
O'Brien, E .
BRITISH MEDICAL JOURNAL, 2000, 320 (7238) :815-816
[10]   Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adults [J].
O'Brien, E ;
Pickering, T ;
Asmar, R ;
Myers, M ;
Parati, G ;
Staessen, J ;
Mengden, T ;
Imai, Y ;
Waeber, B ;
Palatini, P ;
Atkins, N ;
Gerin, W .
BLOOD PRESSURE MONITORING, 2002, 7 (01) :3-17