A luminescent oxygen channeling immunoassay for the determination of insulin in human plasma

被引:66
作者
Poulsen, Fritz [1 ]
Jensen, Kirsten Borres [1 ]
机构
[1] Diabet Res Unit, Bagsvaerd, Denmark
关键词
sandwich; high throughput; biomarker; LOCI (TM); homogenous; chemiluminescent;
D O I
10.1177/1087057106297566
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The authors describe a homogeneous, sensitive, and rapid bead-based sandwich immunoassay with a broad analytical range for quantifying insulin in human plasma. The assay was performed as a 2-step reaction by incubating the sample with a mixture of biotinylated anti-insulin antibody and beads covalently coated with anti-insulin antibody for I h. This was followed by incubation with beads covalently coated with streptavidin for 30 min. After the incubation steps, light generated from a chemiluminescent reaction within the beads was quantitated. The assay was run in 384-well plates with a sample volume of 5 mu L. The analytical range extended from 1 to 10,000 pM. Intra-assay precision (% coefficient of variation) ranged from 1.9% to 3.8% for various insulin concentrations. Interassay precision ranged from 4.6% to 7.3%. Assay detection limit was 0.3 pM. There was no interference from moderate hemolysis (with hemoglobin up to 375 mg/dL), bilirubin (up to at least 50 mg/dL), triglyceride (up to at least 1000 mg/dQ, biotin (up to at least 7.7 ng/mL), or ascorbic acid (up to 100 mg/dL). However, gross hemolysis did affect the assay. Comparable results were obtained for plasma (ethylenediamine tetra-acetic acid, citrate, and heparin treated) and serum. The correlation with enzyme-linked immunosorbent assay (ELISA) was good (y = 1.25x + 1.19, R-2 = 0.98). This method is convenient and represents an alternative to ELISA.
引用
收藏
页码:240 / 247
页数:8
相关论文
共 17 条
[1]  
ANDERSEN L, 1993, CLIN CHEM, V39, P578
[2]   BIOTINYLATION OF HUMAN C-3 [J].
BERGER, M ;
GAITHER, TA ;
COLE, RM ;
CHUSED, TM ;
HAMMER, CH ;
FRANK, MM .
MOLECULAR IMMUNOLOGY, 1982, 19 (07) :857-864
[3]  
BERGGREN PO, 2005, HUPO 4 ANN WORLD C M
[4]   Pharmacokinetic screening for the selection of new drug discovery candidates is greatly enhanced through the use of liquid chromatography atmospheric pressure ionization tandem mass spectrometry [J].
Bryant, MS ;
Korfmacher, WA ;
Wang, SY ;
Nardo, C ;
Nomeir, AA ;
Lin, CC .
JOURNAL OF CHROMATOGRAPHY A, 1997, 777 (01) :61-66
[5]  
CLARET EJ, 2005, SBS 10 ANN C ORL FL
[6]  
GLICK MR, 1986, CLIN CHEM, V32, P470
[7]   A sensitive, robust high-throughput electrochemiluminescence assay for rat insulin [J].
Golla, R ;
Seethala, R .
JOURNAL OF BIOMOLECULAR SCREENING, 2004, 9 (01) :62-70
[8]   Solid-phase immunoassays in mechanism-based drug discovery: Their application in the development of inhibitors of the molecular chaperone heat-shock protein 90 [J].
Hardcastle, A ;
Boxall, K ;
Richards, J ;
Tomlin, P ;
Sharp, S ;
Clarke, P ;
Workman, P ;
Aherne, W .
ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2005, 3 (03) :273-285
[9]   Quantitative screening and matrix effect studies of drug discovery compounds in monkey plasma using fast-gradient liquid chromatography/tandem mass spectrometry [J].
Hsieh, YS ;
Chintala, M ;
Mei, H ;
Agans, J ;
Brisson, JM ;
Ng, K ;
Korfmacher, WA .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2001, 15 (24) :2481-2487
[10]   Multiplexed analysis of biomarkers related to obesity and the metabolic syndrome in human plasma, using the luminex-100 system [J].
Liu, MY ;
Xydakis, AM ;
Hoogeveen, RC ;
Jones, PH ;
Smith, EOB ;
Nelson, KW ;
Ballantyne, CM .
CLINICAL CHEMISTRY, 2005, 51 (07) :1102-1109