The authors describe a homogeneous, sensitive, and rapid bead-based sandwich immunoassay with a broad analytical range for quantifying insulin in human plasma. The assay was performed as a 2-step reaction by incubating the sample with a mixture of biotinylated anti-insulin antibody and beads covalently coated with anti-insulin antibody for I h. This was followed by incubation with beads covalently coated with streptavidin for 30 min. After the incubation steps, light generated from a chemiluminescent reaction within the beads was quantitated. The assay was run in 384-well plates with a sample volume of 5 mu L. The analytical range extended from 1 to 10,000 pM. Intra-assay precision (% coefficient of variation) ranged from 1.9% to 3.8% for various insulin concentrations. Interassay precision ranged from 4.6% to 7.3%. Assay detection limit was 0.3 pM. There was no interference from moderate hemolysis (with hemoglobin up to 375 mg/dL), bilirubin (up to at least 50 mg/dL), triglyceride (up to at least 1000 mg/dQ, biotin (up to at least 7.7 ng/mL), or ascorbic acid (up to 100 mg/dL). However, gross hemolysis did affect the assay. Comparable results were obtained for plasma (ethylenediamine tetra-acetic acid, citrate, and heparin treated) and serum. The correlation with enzyme-linked immunosorbent assay (ELISA) was good (y = 1.25x + 1.19, R-2 = 0.98). This method is convenient and represents an alternative to ELISA.
引用
收藏
页码:240 / 247
页数:8
相关论文
共 17 条
[11]  
Lovendahl P, 2002, J ANIM SCI, V80, P191
[12]  
MYRICK JE, 1989, CLIN CHEM, V35, P37
[13]   Lead development: Validation and application of high throughput screening for determination of pharmacokinetic parameters for enzyme inhibitors [J].
Singh, J ;
Soloweij, J ;
Allen, M ;
Killar, L ;
Ator, M .
BIOORGANIC & MEDICINAL CHEMISTRY, 1996, 4 (05) :639-643
[14]  
TOIVONEN E, 1986, CLIN CHEM, V32, P637
[15]   A "Mix and read" assay for insulin using fluorometric microvolume assay technology [J].
Toney, JH ;
Ogawa, A ;
Blair, M ;
Park, YW .
ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2003, 1 (04) :521-525
[16]   LUMINESCENT OXYGEN CHANNELING IMMUNOASSAY - MEASUREMENT OF PARTICLE BINDING-KINETICS BY CHEMILUMINESCENCE [J].
ULLMAN, EF ;
KIRAKOSSIAN, H ;
SINGH, S ;
WU, ZP ;
IRVIN, BR ;
PEASE, JS ;
SWITCHENKO, AC ;
IRVINE, JD ;
DAFFORN, A ;
SKOLD, CN ;
WAGNER, DB .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1994, 91 (12) :5426-5430
[17]  
Ullman EF, 1996, CLIN CHEM, V42, P1518