Combination of free-breathing and breathhold steady-state free precession magnetic resonance angiography for detection of coronary artery stenoses

被引:30
作者
Dewey, M
Teige, F
Schnapauff, D
Laule, M
Borges, AC
Rutsch, W
Hamm, B
Taupitz, M
机构
[1] Humboldt Univ, Charite Univ Med Berlin, Inst Radiol, D-10117 Berlin, Germany
[2] Free Univ Berlin, Charite Med Sch, Dept Radiol, D-1000 Berlin, Germany
[3] Free Univ Berlin, Charite Med Sch, Dept Cardiol, D-1000 Berlin, Germany
关键词
coronary disease; diagnosis; magnetic resonance imaging; coronary vessels; angiography;
D O I
10.1002/jmri.20568
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To analyze the incremental diagnostic value of a combination of two approaches (free-breathing and breath-hold) vs. the sole free-breathing approach to coronary magnetic resonance angiography (CMRA) for detection of significant stenoses. Materials and Methods: Thirty patients were consecutively included in this prospective trial. CMRA was performed on a 1.5-T MR scanner (Magnetom Sonata, Siemens) using a balanced steady-state free precession (SSFP) sequence during free-breathing (2.4 x 0.9 x 0.7 mm(3)). Breathholding acquisitions (3.0 x 1.5 x 0.7 mm(3)) were only performed in cases in which the quality of free-breatbing CMRA precluded assessment. Patients with contraindications to CMRA, claustrophobia, or nonassessable images were not excluded from the assessment of diagnostic accuracy (intention-to-diagnose design). Results: In 60% of all free-breathing coronary acquisitions the image quality was adequate for diagnostic assessment. For the remaining 40% of the cases, breathhold acquisitions were obtained. The sensitivity, specificity, nonassessable rate, and accuracy in identifying main coronary branches with significant stenoses using the combination of both breathing approaches and the free-breathing approach alone were 65% vs. 32%, 73% vs. 53%, 24% vs. 52%, and 71 % vs. 46%, respectively (P < 0.001). Conclusion: In this consecutive cohort of patients. the combination of free-breathing and breathhold CMRA significantly improved diagnostic accuracy. Nevertheless, even this combination did not reach accuracies sufficient for routine clinical application.
引用
收藏
页码:674 / 681
页数:8
相关论文
共 28 条
[1]   Comparison of image quality in contrast-enhanced coronary artery visualization by electron beam tomography and retrospectively electrocardiogram-gated multislice spiral computed tomography [J].
Achenbach, S ;
Giesler, T ;
Ropers, D ;
Ulzheimer, S ;
Anders, K ;
Wenkel, E ;
Pohle, K ;
Kachelriess, M ;
Derlien, H ;
Kalender, WA ;
Daniel, WG ;
Bautz, W ;
Baum, U .
INVESTIGATIVE RADIOLOGY, 2003, 38 (02) :119-128
[2]  
Austen W G, 1975, Circulation, V51, P5
[3]   Coronary artery imaging with real-time navigator three-dimensional turbo-field-echo MR coronary angiography: Initial experience [J].
Bogaert, J ;
Kuzo, R ;
Dymarkowski, S ;
Beckers, R ;
Piessens, J ;
Rademakers, FE .
RADIOLOGY, 2003, 226 (03) :707-716
[4]   Towards complete and accurate reporting of studies of diagnostic accuracy: The STARD initiative [J].
Bossuyt, PM ;
Reitsma, JB ;
Bruns, DE ;
Gatsonis, CA ;
Glasziou, PP ;
Irwig, LM ;
Lijmer, JG ;
Moher, D ;
Rennie, D ;
de Vet, HCW .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (01) :51-56
[5]   Improved coronary artery definition with T2-weighted, free-breathing, three-dimensional coronary MRA [J].
Botnar, RM ;
Stuber, M ;
Danias, PG ;
Kissinger, KV ;
Manning, WJ .
CIRCULATION, 1999, 99 (24) :3139-3148
[6]   Clinical utility of computed tomography and magnetic resonance techniques for noninvasive coronary angiography [J].
Budoff, MJ ;
Achenbach, S ;
Duerinckx, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (11) :1867-1878
[7]   Relationship between motion of coronary arteries and diaphragm during free breathing: Lessons from real-time MR imaging [J].
Danias, PG ;
Stuber, M ;
Botnar, RM ;
Kissinger, KV ;
Edelman, RR ;
Manning, WJ .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 172 (04) :1061-1065
[8]   Breath-hold three-dimensional true-FISP imaging of coronary arteries using asymmetric sampling [J].
Deshpande, VS ;
Shea, SM ;
Chung, YC ;
McCarthy, RM ;
Finn, JP ;
Li, DB .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 15 (04) :473-478
[9]   3D magnetization-prepared true-FISP: A new technique for imaging coronary arteries [J].
Deshpande, VS ;
Shea, SM ;
Laub, G ;
Simonetti, OP ;
Finn, JP ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) :494-502
[10]   Multisegment and halfscan reconstruction of 16-slice computed tomography for detection of coronary artery stenoses [J].
Dewey, M ;
Laule, M ;
Krug, L ;
Schnapauff, D ;
Rogalla, P ;
Rutsch, WF ;
Hamm, B ;
Lembcke, A .
INVESTIGATIVE RADIOLOGY, 2004, 39 (04) :223-229