Biosensor arrays for simultaneous measurement of glucose, lactate, glutamate, and glutamine

被引:146
作者
Moser, I [1 ]
Jobst, G [1 ]
Urban, GA [1 ]
机构
[1] Univ Freiburg, IMTEK, D-79110 Freiburg, Germany
关键词
BioMEMS; monitoring; integration; microfluidics; cross talk free; stability;
D O I
10.1016/S0956-5663(01)00298-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
For simultaneous measurement of glucose, lactate, glutamine, and glutamate a biosensor array is implemented in a micro flow-system thus giving a microsystem. The microsystem consists of a glass chip with the integrated biosensor array and a bottom part, which comprises a gold counter electrode, a 300 m thick seal, and electrical interconnection lines. The flow device has a total internal volume of 2.1 or 6 mul when integrated with a mixer on chip. The biosensors with no crosstalking and high long term stability were produced by modifying the electrochemical transducers and utilizing photopatternable enzyme membranes. The use of appropriate miniaturization technology leads to mass producable devices for in vivo and ex vivo applications in whole blood and fermentation broth. Due to a novel glutaminase with an activity optimum in the neutral pH range direct and simultaneous monitoring of glutamine together with glucose, lactate, and glutamate could be performed. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:297 / 302
页数:6
相关论文
共 10 条
[1]   Mass producible miniaturized flow through a device with a biosensor array [J].
Jobst, G ;
Moser, I ;
Svasek, P ;
Varahram, M ;
Trajanoski, Z ;
Wach, P ;
Kotanko, P ;
Skrabal, F ;
Urban, G .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 43 (1-3) :121-125
[2]   Thin-film microbiosensors for glucose-lactate monitoring [J].
Jobst, G ;
Moser, I ;
Varahram, M ;
Svasek, P ;
Aschauer, E ;
Trajanoski, Z ;
Wach, P ;
Kotanko, P ;
Skrabal, F ;
Urban, G .
ANALYTICAL CHEMISTRY, 1996, 68 (18) :3173-3179
[3]   MINIATURIZED THIN-FILM GLUTAMATE AND GLUTAMINE BIOSENSORS [J].
MOSER, I ;
JOBST, G ;
ASCHAUER, E ;
SVASEK, P ;
VARAHRAM, M ;
URBAN, G ;
ZANIN, VA ;
TJOUTRINA, GY ;
ZHARIKOVA, AV ;
BEREZOV, TT .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (6-7) :527-532
[4]  
MOSER I, 1998, P MUTAS 98 BANFF ALT
[5]   Plastic film carbon electrodes: Enzymatic modification for on line, continuous, and simultaneous measurement of lactate and glucose using microdialysis sampling [J].
Osborne, PG ;
Niwa, O ;
Yamamoto, K .
ANALYTICAL CHEMISTRY, 1998, 70 (09) :1701-1706
[6]   Simultaneous monitoring of glucose and lactate by an interference and cross-talk free dual electrode amperometric biosensor based on electropolymerized thin films [J].
Palmisano, F ;
Rizzi, R ;
Centonze, D ;
Zambonin, PG .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (9-10) :531-539
[7]  
PECHOV AA, 1985, B EXP BIOL MED, V99, P557
[8]   Miniaturized real-time monitoring system for L-lactate and glucose using microfabricated multi-enzyme sensors [J].
Perdomo, J ;
Hinkers, H ;
Sundermeier, C ;
Seifert, W ;
Morell, OM ;
Knoll, M .
BIOSENSORS & BIOELECTRONICS, 2000, 15 (9-10) :515-522
[9]  
SANDGODKAR H, 1996, ANAL CHEM, V68, P779
[10]   Chemical cross-talk in flow-type integrated enzyme sensors [J].
Suzuki, M ;