Radiation Dose Reduction in Abdominal Computed Tomography During the Late Hepatic Arterial Phase Using a Model-Based Iterative Reconstruction Algorithm How Low Can We Go?

被引:61
作者
Husarik, Daniela B. [2 ]
Marin, Daniele [1 ]
Samei, Ehsan [1 ]
Richard, Samuel [1 ]
Chen, Baiyu [1 ]
Jaffe, Tracy A. [1 ]
Bashir, Mustafa R. [1 ]
Nelson, Rendon C. [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Univ Zurich Hosp, CH-8091 Zurich, Switzerland
基金
瑞士国家科学基金会;
关键词
computed tomography; iterative reconstruction; raw data; radiation dose; image quality; FILTERED BACK-PROJECTION; IMPROVED IMAGE QUALITY; CT ANGIOGRAPHY; CHEST; ROUTINE; NOISE;
D O I
10.1097/RLI.0b013e318251eafd
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Objective: The aim of this study was to compare the image quality of abdominal computed tomography scans in an anthropomorphic phantom acquired at different radiation dose levels where each raw data set is reconstructed with both a standard convolution filtered back projection (FBP) and a full model-based iterative reconstruction (MBIR) algorithm. Materials and Methods: An anthropomorphic phantom in 3 sizes was used with a custom-built liver insert simulating late hepatic arterial enhancement and containing hypervascular liver lesions of various sizes. Imaging was performed on a 64-section multidetector-row computed tomography scanner (Discovery CT750 HD; GE Healthcare, Waukesha, WI) at 3 different tube voltages for each patient size and 5 incrementally decreasing tube current-time products for each tube voltage. Quantitative analysis consisted of contrast-to-noise ratio calculations and image noise assessment. Qualitative image analysis was performed by 3 independent radiologists rating subjective image quality and lesion conspicuity. Results: Contrast-to-noise ratio was significantly higher and mean image noise was significantly lower on MBIR images than on FBP images in all patient sizes, at all tube voltage settings, and all radiation dose levels (P < 0.05). Overall image quality and lesion conspicuity were rated higher for MBIR images compared with FBP images at all radiation dose levels. Image quality and lesion conspicuity on 25% to 50% dose MBIR images were rated equal to full-dose FBP images. Conclusion: This phantom study suggests that depending on patient size, clinically acceptable image quality of the liver in the late hepatic arterial phase can be achieved with MBIR at approximately 50% lower radiation dose compared with FBP.
引用
收藏
页码:468 / 474
页数:7
相关论文
共 41 条
[1]
Iterative reconstruction in image space (IRIS) in cardiac computed tomography: initial experience [J].
Bittencourt, Marcio Sommer ;
Schmidt, Bernhard ;
Seltmann, Martin ;
Muschiol, Gerd ;
Ropers, Dieter ;
Daniel, Werner Guenther ;
Achenbach, Stephan .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2011, 27 (07) :1081-1087
[2]
Dynamic Contrast-Enhanced CT Studies Balancing Patient Exposure and Image Noise [J].
Brix, Gunnar ;
Lechel, Ursula ;
Petersheim, Markus ;
Krissak, Radko ;
Fink, Christian .
INVESTIGATIVE RADIOLOGY, 2011, 46 (01) :64-70
[3]
THEORY OF IMAGE RECONSTRUCTION IN COMPUTED TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G .
RADIOLOGY, 1975, 117 (03) :561-572
[4]
A New Iodinated Liver Phantom for the Quantitative Evaluation of Advanced CT Acquisition and Reconstruction Techniques [J].
Chen, Baiyu ;
Marin, Daniele ;
Samei, Ehsan .
MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
[5]
Impact of Adaptive Statistical Iterative Reconstruction (ASIR) on Radiation Dose and Image Quality in Aortic Dissection Studies: A Qualitative and Quantitative Analysis [J].
Cornfeld, Daniel ;
Israel, Gary ;
Detroy, Ezra ;
Bokhari, Jamal ;
Mojibian, Hamid .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (03) :W336-W340
[6]
Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[7]
Reducing the Radiation Dose for CT Colonography Using Adaptive Statistical Iterative Reconstruction: A Pilot Study [J].
Flicek, Kristina T. ;
Hara, Amy K. ;
Silva, Alvin C. ;
Wu, Qing ;
Peter, Mary B. ;
Johnson, C. Daniel .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (01) :126-131
[8]
CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT [J].
Gervaise, Alban ;
Osemont, Benoit ;
Lecocq, Sophie ;
Noel, Alain ;
Micard, Emilien ;
Felblinger, Jacques ;
Blum, Alain .
EUROPEAN RADIOLOGY, 2012, 22 (02) :295-301
[9]
CT iterative reconstruction in image space: A phantom study [J].
Ghetti, C. ;
Ortenzia, O. ;
Serreli, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :161-165
[10]
Habets J, 2011, INT J CARDIOVASC IMA