Dynamic Contrast-Enhanced CT Studies Balancing Patient Exposure and Image Noise

被引:34
作者
Brix, Gunnar [1 ]
Lechel, Ursula [1 ]
Petersheim, Markus [2 ]
Krissak, Radko [3 ]
Fink, Christian [3 ]
机构
[1] Bundesamt Strahlenschutz, Abt Med & Beruflichen Strahlenschutz, D-85764 Oberschleissheim, Germany
[2] Univ Heidelberg, Dept Radiat Oncol, Univ Med Ctr Mannheim, D-6800 Mannheim, Germany
[3] Univ Heidelberg, Inst Clin Radiol & Nucl Med, D-6800 Mannheim, Germany
关键词
dynamic contrast-enhanced CT; patient exposure; image noise; spatial resolution; biologic radiation effects; CEREBRAL BLOOD-FLOW; COMPUTED-TOMOGRAPHY; PERFUSION; BRAIN; EXCHANGE; MODEL;
D O I
10.1097/RLI.0b013e3181f33b35
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To establish the essential basis for balancing the dose versus noise trade-off in dynamic contrast-enhanced (DCE) CT by means of a phantom study. Materials and Methods: Measurements were performed at a 64-section dual-source system, using the default protocols for DCE imaging (40 scans) of the trunk (current-time product per scan, 100 mAs; voltage, 120 kVp; pixel size, 0.9 x 0.9 x 8 mm(3); CTDIvol per examination, 264 mGy) and head (270 mAs, 80 kVp, 0.45 x 0.45 x 8 mm(3), 429 mGy). For 3 representative sections of an anthropomorphic phantom (head, upper abdomen, pelvis) transaxial dose distributions were measured by thermoluminescent dosimeters. The image noise was determined for 5 values of the current-time product (but otherwise identical parameter settings) and 4 pixel resolutions at a water-filled trunk and head phantom. Results: Highest exposures occurred at the periphery of the trunk and head with maximum skin entrance doses of about 300 mGy. Effective doses related to the 3 exposure scenarios were between 4 and 20 mSv, but were not at all predictive of local exposure levels. The image noise was inversely proportional to the square root of the current-time product and, with restrictions, to the pixel size. Noise levels determined for the standard settings were 13.8 HU (trunk) and 4.4 HU (head) and thus comparable with the contrast enhancement typically detected in tumors and ischemic brain tissues, respectively. Conclusions: The opposing requirements of risk and noise limitation in DCE-CT cannot be balanced without substantially reducing the spatial resolution. But even so, local radiation exposures are rather high for a diagnostic procedure. Indications to perform a DCE examination should thus be strictly limited to patients who really benefit from it.
引用
收藏
页码:64 / 70
页数:7
相关论文
共 24 条
  • [1] Radiation Cataractogenesis: A Review of Recent Studies
    Ainsbury, E. A.
    Bouffler, S. D.
    Doerr, W.
    Graw, J.
    Muirhead, C. R.
    Edwards, A. A.
    Cooper, J.
    [J]. RADIATION RESEARCH, 2009, 172 (01) : 1 - 9
  • [2] [Anonymous], 2007, The 2007 Recommendations of the International Commission on Radiological Protection
  • [4] Assessment of a theoretical formalism for dose estimation in CT: an anthropomorphic phantom study
    Brix, G
    Lechel, U
    Veit, R
    Truckenbrodt, R
    Stamm, G
    Coppenrath, EM
    Griebel, J
    Nagel, HD
    [J]. EUROPEAN RADIOLOGY, 2004, 14 (07) : 1275 - 1284
  • [5] Regional blood flow capillary permeability, and compartmental volumes: Measurement with dynamic CT - Initial experience
    Brix, G
    Bahner, ML
    Hoffmann, UJ
    Horvath, A
    Schreiber, W
    [J]. RADIOLOGY, 1999, 210 (01) : 269 - 276
  • [6] Estimation of tissue perfusion by dynamic contrast-enhanced imaging: simulation-based evaluation of the steepest slope method
    Brix, Gunnar
    Zwick, Stefan
    Griebel, Juergen
    Fink, Christian
    Kiessling, Fabian
    [J]. EUROPEAN RADIOLOGY, 2010, 20 (09) : 2166 - 2175
  • [7] Pharmacokinetic analysis of tissue microcirculation using nested models: Multimodel inference and parameter identifiability
    Brix, Gunnar
    Zwick, Stefan
    Kiessling, Fabian
    Griebel, Juergen
    [J]. MEDICAL PHYSICS, 2009, 36 (07) : 2923 - 2933
  • [8] BROOKS R A, 1976, Medical Physics (Woodbury), V3, P237, DOI 10.1118/1.594240
  • [9] European Commission, 2000, 19604 EUR EN EUR COM
  • [10] X-ray imaging and the skin: Radiation biology, patient dosimetry and observed effects
    Geleijns, J
    Wondergem, J
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 121 - 125