A novel 675.2 nm diode laser densitometer for use with GafChromic films

被引:14
作者
Ertl, A
Kellermann, PO
Zehetmayer, M
Schöggl, A
Kindl, P
Maitz, AH
机构
[1] Univ Vienna, Dept Neurosurg, A-1090 Vienna, Austria
[2] Tech Univ Vienna, Inst Solid State Elect, Vienna, Austria
[3] Univ Vienna, Dept Ophthalmol, A-1010 Vienna, Austria
[4] Graz Tech Univ, Inst Tech Phys, Graz, Austria
[5] Univ Pittsburgh, Med Ctr, Dept Neurol Surg, Pittsburgh, PA USA
关键词
radiochromic film; densitometer; dosimetry; GafChromic film; gamma knife;
D O I
10.1118/1.598592
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this article we compare the accuracy of a diode laser densitometer emitting 675.2 nm to that of a commercial He-Ne laser densitometer emitting 632.8 nm for GafChromic MD-55 film readout. A Leksell gamma unit (AB Elekta Stockholm, Sweden) Model B with a 14 and 8 mm collimator at the same isocenter (combined 11 mm collimator) was used to irradiate GafChromic MD-55 films. Dose response curves, dose cross profile and FWHM were measured with a custom-designed diode laser scanning device, emitting light at 675.2 nm. The same data were recorded with a commercial He-Ne laser densitometer (PTW FIPS Plus, Freiburg, Germany), emitting light at 632.8 nm. Both measurements were compared to dose cross profiles of a radiosurgery dose planning program (GammaPlan 5.12, Elekta, Sweden). Compared to the commercial He-Ne laser densitometer, the custom-designed diode laser scanning device showed better agreement with the calculated dose cross profile. For two axes, the full width half maxima (FWHM) of the diode laser scanning device was within 0.1 mm deviation compared to the data calculated by the dose planning program. The FWHM of the commercial He-Ne laser densitometer was less accurate (1.6 and 2.1 mm deviation). Our data show that a diode laser scanning device using a Light source emitting 675.2 nm increases the accuracy of a GafChromic MD-55 film readout. This greater accuracy may be related to the diode laser measuring the optical density close to maximum absorption of the GafChromic film MD-55 (671-675 nm). (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)00305-3].
引用
收藏
页码:834 / 838
页数:5
相关论文
共 18 条
[1]   Monte Carlo calculation of single-beam dose profiles used in a gamma knife treatment planning system [J].
Cheung, JYC ;
Yu, KN ;
Yu, CP ;
Ho, RTK .
MEDICAL PHYSICS, 1998, 25 (09) :1673-1675
[2]  
CHU RDH, 1990, RADIAT PHYS CHEM, V35, P767
[3]   Shuttle dose at the Vienna Leksell Gamma Knife [J].
Ertl, A ;
Zehetmayer, M ;
Schoggl, A ;
Kitz, K ;
Koschuch, R ;
Stadtmann, H ;
Kindl, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1567-1578
[4]   A new method of readout in radiochromic film dosimetry [J].
Kellermann, PO ;
Ertl, A ;
Gornik, E .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2251-2263
[5]   Single-mode and single-beam surface emission from visible red GaInP/AlGaInP laser diodes [J].
Kellermann, PO ;
Golshani, A ;
Kock, A ;
Gornik, E ;
Gauggel, HP ;
Winterhoff, R ;
Pilkuhn, MH .
APPLIED PHYSICS LETTERS, 1997, 70 (18) :2374-2376
[6]   GafChromic MD-55: Investigated as a precision dosimeter [J].
Klassen, NV ;
van der Zwan, L ;
Cygler, J .
MEDICAL PHYSICS, 1997, 24 (12) :1924-1934
[7]   THE USE OF A RADIOCHROMIC DETECTOR FOR THE DETERMINATION OF STEREOTAXIC RADIOSURGERY DOSE CHARACTERISTICS [J].
MCLAUGHLIN, WL ;
SOARES, CG ;
SAYEG, JA ;
MCCULLOUGH, EC ;
KLINE, RW ;
WU, A ;
MAITZ, AH .
MEDICAL PHYSICS, 1994, 21 (03) :379-388
[8]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[9]   Dosimetric characteristics of an improved radiochromic film [J].
Meigooni, AS ;
Sanders, MF ;
Ibbott, GS ;
Szeglin, SR .
MEDICAL PHYSICS, 1996, 23 (11) :1883-1888
[10]  
MEIGOONI AS, 1993, MED PHYS, V20, P1298