Investigation of in vivo 6DOF total knee arthoplasty kinematics using a dual orthogonal fluoroscopic system

被引:68
作者
Hanson, GR
Suggs, JF
Freiberg, AA
Durbhakula, S
Li, GA
机构
[1] Massachusetts Gen Hosp, Dept Orthopaed Surg, Bioengn Lab, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA 02114 USA
[3] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
关键词
in vivo kinematics; biplanar fluoroscopy; total knee arthoplasty (TKA); image analysis; knee;
D O I
10.1002/jor.20141
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Fluoroscopic techniques have been recently used to detect in vivo knee joint kinematics. This article presents a technique that uses two fluoroscopes to form a dual orthogonal fluoroscopic system for accurately measuring in vivo 6DOF total knee arthoplasty (TKA) kinematics. The system was rigorously validated and used to investigate in vivo kinematics of 12 patients after cruciate-retaining TKA. In a repeatability study, the pose of two different TKA components was reproduced with standard deviations (SD) of 0.17 mm and 0.57 degrees about all three axes. In an accuracy study, the reproduced component positions were compared to the known component positions. Position and rotation mean errors were all within 0.11 mm and 0.24 degrees, with SD within 0.11 mm and 0.48 degrees, respectively. The results of this study show that the matching process of the imaging system is able to accurately reproduce the spatial positions and orientations of both the femoral and tibial components. For CR TKA patients, a consistent anterior femoral translation was observed with flexion through 45 degrees of flexion, and thereafter, the femur translated posteriorly with further flexion. The medial-lateral translation was measured to be less than 2 mm throughout the entire flexion range. Internal tibial rotation steadily increased through maximum flexion by approximately 6 degrees. Varus rotation was also measured with flexion but had a mean magnitude less than 2.0 degrees. In conclusion, the dual orthogonal fluoroscopic system accurately detects TKA kinematics and is applicable towards otherjoints of the musculoskeletal system, including the wrist, elbow, shoulder, ankle, and spine. (c) 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 24:974-981, 2006.
引用
收藏
页码:974 / 981
页数:8
相关论文
共 17 条
[1]   The use of functional analysis in evaluating knee kinematics [J].
Andriacchi, TP ;
Dyrby, CO ;
Johnson, TS .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :44-53
[2]   In vivo three-dimensional knee kinematics using a biplanar image-matching technique [J].
Asano, T ;
Akagi, M ;
Tanaka, K ;
Tamura, J ;
Nakamura, T .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (388) :157-166
[3]   Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy [J].
Banks, SA ;
Hodge, WA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (06) :638-649
[4]   Multicenter determination of in vivo kinematics after total knee arthroplasty [J].
Dennis, DA ;
Komistek, RD ;
Mahfouz, MR ;
Haas, BD ;
Stiehl, JB .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (416) :37-57
[5]  
FREGLY BJ, 2004, IN PRESS TRANS ORTHO
[6]   A new model-based RSA method validated using CAD models and models from reversed engineering [J].
Kaptein, BL ;
Valstar, ER ;
Stoel, BC ;
Rozing, PM ;
Reiber, JHC .
JOURNAL OF BIOMECHANICS, 2003, 36 (06) :873-882
[7]   In vivo fluoroscopic analysis of the normal human knee [J].
Komistek, RD ;
Dennis, DA ;
Mahfouz, M .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :69-81
[8]   RECOVERING THE POSITION AND ORIENTATION OF FREE-FORM OBJECTS FROM IMAGE CONTOURS USING 3D DISTANCE MAPS [J].
LAVALLEE, S ;
SZELISKI, R .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1995, 17 (04) :378-390
[9]   In vivo elongation of the anterior cruciate ligament and posterior cruciate ligament during knee flexion [J].
Li, G ;
DeFrate, LE ;
Sun, H ;
Gill, TJ .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2004, 32 (06) :1415-1420
[10]   Cruciate-retaining and cruciate-substituting total knee arthroplasty -: An in vitro comparison of the kinematics under muscle loads [J].
Li, G ;
Zayontz, S ;
Most, E ;
Otterberg, E ;
Sabbag, K ;
Rubash, HE .
JOURNAL OF ARTHROPLASTY, 2001, 16 (08) :150-156