Investigation of in vivo 6DOF total knee arthoplasty kinematics using a dual orthogonal fluoroscopic system

被引:68
作者
Hanson, GR
Suggs, JF
Freiberg, AA
Durbhakula, S
Li, GA
机构
[1] Massachusetts Gen Hosp, Dept Orthopaed Surg, Bioengn Lab, Boston, MA 02114 USA
[2] Harvard Univ, Sch Med, Boston, MA 02114 USA
[3] MIT, Dept Mech Engn, Cambridge, MA 02139 USA
关键词
in vivo kinematics; biplanar fluoroscopy; total knee arthoplasty (TKA); image analysis; knee;
D O I
10.1002/jor.20141
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Fluoroscopic techniques have been recently used to detect in vivo knee joint kinematics. This article presents a technique that uses two fluoroscopes to form a dual orthogonal fluoroscopic system for accurately measuring in vivo 6DOF total knee arthoplasty (TKA) kinematics. The system was rigorously validated and used to investigate in vivo kinematics of 12 patients after cruciate-retaining TKA. In a repeatability study, the pose of two different TKA components was reproduced with standard deviations (SD) of 0.17 mm and 0.57 degrees about all three axes. In an accuracy study, the reproduced component positions were compared to the known component positions. Position and rotation mean errors were all within 0.11 mm and 0.24 degrees, with SD within 0.11 mm and 0.48 degrees, respectively. The results of this study show that the matching process of the imaging system is able to accurately reproduce the spatial positions and orientations of both the femoral and tibial components. For CR TKA patients, a consistent anterior femoral translation was observed with flexion through 45 degrees of flexion, and thereafter, the femur translated posteriorly with further flexion. The medial-lateral translation was measured to be less than 2 mm throughout the entire flexion range. Internal tibial rotation steadily increased through maximum flexion by approximately 6 degrees. Varus rotation was also measured with flexion but had a mean magnitude less than 2.0 degrees. In conclusion, the dual orthogonal fluoroscopic system accurately detects TKA kinematics and is applicable towards otherjoints of the musculoskeletal system, including the wrist, elbow, shoulder, ankle, and spine. (c) 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 24:974-981, 2006.
引用
收藏
页码:974 / 981
页数:8
相关论文
共 17 条
[11]   Feasibility of using orthogonal fluoroscopic images to measure in vivo joint kinematics [J].
Li, GA ;
Wuerz, TH ;
DeFrate, LE .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2004, 126 (02) :314-318
[12]   A robust method for registration of three-dimensional knee implant models to two-dimensional fluoroscopy images [J].
Mahfouz, MR ;
Hoff, WA ;
Komistek, RD ;
Dennis, DA .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2003, 22 (12) :1561-1574
[13]   Femoral rollback after cruciate-retaining and stabilizing total knee arthroplasty [J].
Most, E ;
Zayontz, S ;
Li, GA ;
Otterberg, E ;
Sabbag, K ;
Rubash, HE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (410) :101-113
[14]   A comparison of similarity measures for use in 2-D-3-D medical image registration [J].
Penney, GP ;
Weese, J ;
Little, JA ;
Desmedt, P ;
Hill, DLG ;
Hawkes, DJ .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (04) :586-595
[15]   In vivo kinematics of mobile-bearing knee arthroplasty in deep knee bending motion [J].
Watanabe, T ;
Yamazaki, T ;
Sugamoto, K ;
Tomita, T ;
Hashimoto, H ;
Maeda, D ;
Tamura, S ;
Ochi, T ;
Yoshikawa, H .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (05) :1044-1049
[16]   In vivo measurement of 3-D skeletal kinematics from sequences of biplane radiographs:: Application to knee kinematics [J].
You, BM ;
Siy, P ;
Anderst, W ;
Tashman, S .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (06) :514-525
[17]   A model-based method for the reconstruction of total knee replacement kinematics [J].
Zuffi, S ;
Leardini, A ;
Catani, F ;
Fantozzi, S ;
Cappello, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1999, 18 (10) :981-991