Accuracy of proximal caries depth measurements:: Comparison between limited cone beam computed tomography, storage phosphor and film radiography

被引:52
作者
Akdeniz, B. Guniz [1 ]
Grondahl, Hans-Goran
Magnusson, Bengt
机构
[1] Ege Univ, Sch Dent, Dishekimligi Fak Oral Diagnoz & Rad AD, Dept Oral Diag & Radiol, TR-35100 Izmir, Turkey
[2] Univ Gothenburg, Sahlgrenska Acad, Fac Odontol, Dept Oral & Maxillofacial Radiol, Gothenburg, Sweden
[3] Univ Gothenburg, Sahlgrenska Acad, Dept Oral Pathol, Gothenburg, Sweden
关键词
cone beam X-ray CT; digital radiography;
D O I
10.1159/000092226
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The aim of this study was to compare the accuracy of limited cone beam computed tomography (LCBCT), an image plate system and F-speed film in assessing the depth of proximal carious lesions. Radiographs of a dry mandible with sound and carious teeth were obtained with all three methods. In 41 molar and premolar proximal surfaces, 2 observers independently measured lesion depth on the images from the three modalities. The correlation of measurements was assessed with Pearson's correlation analysis. Results from imaging modalities and histological sectioning (gold standard) were compared using Bland-Altman plots. Overall comparison of the depth measurements from the imaging modalities and the gold standard was done using repeated-measures ANOVA. Pairwise comparisons of systems were done by the Bonferroni t test. The correlation between the measurements of the two observers was 0.977 for film, 0.997 for image plate system and 0.998 for LCBCT. Bland-Altman plots revealed that LCBCT agreed very closely with the gold standard while the agreement between the latter and the image plate - or film - images was moderate. The mean difference and 95% limits of agreement between LCBCT and the gold standard were smaller than those between either image plate or F-speed film and the gold standard. The LCBCT method appears as a promising tool for detection and monitoring of proximal carious lesions.
引用
收藏
页码:202 / 207
页数:6
相关论文
共 18 条
[11]   Relationship between histological and radiographic caries lesion depth measured in images from four digital radiography systems [J].
Jacobsen, JH ;
Hansen, B ;
Wenzel, A ;
Hintze, H .
CARIES RESEARCH, 2004, 38 (01) :34-38
[12]   A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results [J].
Mozzo, P ;
Procacci, C ;
Tacconi, A ;
Martini, PT ;
Andreis, IAB .
EUROPEAN RADIOLOGY, 1998, 8 (09) :1558-1564
[13]   INVITRO STUDY INTO THE INFLUENCE OF X-RAY-BEAM ANGULATION ON THE DETECTION OF ARTIFICIAL CARIES DEFECTS ON BITEWING RADIOGRAPHS [J].
VANDERSTELT, PF ;
RUTTIMAN, UE ;
WEBBER, RL ;
HEEMSTRA, P .
CARIES RESEARCH, 1989, 23 (05) :334-341
[14]   In vivo study of approximal caries depth on storage phosphor plate images compared with dental x-ray film [J].
Versteeg, KH ;
Sanderink, GCH ;
Velders, XL ;
vanGinkel, FC ;
vanderStelt, PF .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1997, 84 (02) :210-213
[15]   Digital radiography and caries diagnosis [J].
Wenzel, A .
DENTOMAXILLOFACIAL RADIOLOGY, 1998, 27 (01) :3-11
[16]   Accuracy of computer-automated caries detection in digital radiographs compared with human observers [J].
Wenzel, A ;
Hintze, H ;
Kold, LM ;
Kold, S .
EUROPEAN JOURNAL OF ORAL SCIENCES, 2002, 110 (03) :199-203
[17]   ACCURACY OF CARIES DIAGNOSIS IN DIGITAL IMAGES FROM CHARGE-COUPLED-DEVICE AND STORAGE PHOSPHOR SYSTEMS - AN IN-VITRO STUDY [J].
WENZEL, A ;
BORG, E ;
HINTZE, H ;
GRONDAHL, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (04) :250-254
[18]  
Wenzel A, 2001, CARIES RES, V35, P12