Conceptual design of a high-sensitivity small animal PET camera with 4π coverage

被引:37
作者
Huber, JS [1 ]
Moses, WW [1 ]
机构
[1] Univ Calif Berkeley, Lawrence Berkeley Lab, Berkeley, CA 94720 USA
基金
美国国家卫生研究院;
关键词
D O I
10.1109/23.775569
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present a conceptual design of a high-sensitivity PET camera that completely encloses a small animal in a rectangular volume formed by 6 planar banks of detector modules. The 4 pi geometry and 3 attenuation-length fast scintillators provide significantly higher sensitivity than contemporary animal PET cameras, while the depth of interaction (DOI) measurement and small crystal width achieve isotropic, high spatial resolution. The absolute sensitivity is 24 kcps/mu Ci, similar to 120 times higher than contemporary systems; the true event count rate is increased by covering 10 times the solid angle using 80% efficient detectors. For a 29 g "mouse", the total scatter event rate is 11% of the total true event rate. A short (2 nsec) coincidence window and the absence of out of field activity implicit with whole animal coverage yield a small random fraction. Assuming a maximum system count rate of 10 Mcps (achievable with electronics under development), the noise equivalent count rate as a function of activity concentration has a maximum of 6.6 Mcps at 25 mu Ci/cc. 2D reconstruction algorithms indicate a spatial resolution of 2.3 mm fwhm through most of the field of view.
引用
收藏
页码:498 / 502
页数:5
相关论文
共 20 条
  • [1] THE DESIGN AND PHYSICAL CHARACTERISTICS OF A SMALL ANIMAL POSITRON EMISSION TOMOGRAPH
    BLOOMFIELD, PM
    RAJESWARAN, S
    SPINKS, TJ
    HUME, SP
    MYERS, R
    ASHWORTH, S
    CLIFFORD, KM
    JONES, WF
    BYARS, LG
    YOUNG, J
    ANDREACO, M
    WILLIAMS, CW
    LAMMERTSMA, AA
    JONES, T
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) : 1105 - 1126
  • [2] Design and physical characteristics of a small animal PET using BaF2 crystals and a photosensitive wire chamber
    Bruyndonckx, P
    Liu, X
    Rajeswaran, S
    Smolik, W
    Tavernier, S
    Zhang, SP
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 382 (03) : 589 - 600
  • [3] MicroPET: A high resolution PET scanner for imaging small animals
    Cherry, SR
    Shao, Y
    Silverman, RW
    Meadors, K
    Siegel, S
    Chatziioannou, A
    Young, JW
    Jones, WF
    Moyers, JC
    Newport, D
    Boutefnouchet, A
    Farquhar, TH
    Andreaco, M
    Paulus, MJ
    Binkley, DM
    Nutt, R
    Phelps, ME
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1161 - 1166
  • [4] FULLY 3-DIMENSIONAL POSITRON EMISSION TOMOGRAPHY
    COLSHER, JG
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01) : 103 - 115
  • [5] Use of a YRP:Ce matrix coupled to a position-sensitive photomultiplier for high resolution position emission tomography
    DelGuerra, A
    deNotaristefani, F
    DiDomenico, G
    Giganti, M
    Pani, R
    Piffanelli, A
    Zavattini, G
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) : 1958 - 1962
  • [6] A small animal PET prototype based on LSO crystals read out by avalanche photodiodes
    Fries, O
    Bradbury, SM
    Gebauer, J
    Holl, I
    Lorenz, E
    Renker, D
    Ziegler, SI
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 387 (1-2) : 220 - 224
  • [7] Characterization of a 64 channel PET detector using photodiodes for crystal identification
    Huber, JS
    Moses, WW
    Derenzo, SE
    Ho, MH
    Andreaco, MS
    Paulus, MJ
    Nutt, R
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) : 1197 - 1201
  • [8] Calibration of a PET detector module that measures depth of interaction
    Huber, JS
    Moses, WW
    Virador, PRG
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 1268 - 1272
  • [9] Positron emission tomography (PET) methodology for small animals and its application in radiopharmaceutical preclinical investigation
    Hume, SP
    Jones, T
    [J]. NUCLEAR MEDICINE AND BIOLOGY, 1998, 25 (08) : 729 - 732
  • [10] Pharmacological constraints associated with positron emission tomographic scanning of small laboratory animals
    Hume, SP
    Gunn, RN
    Jones, T
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (02) : 173 - 176