Comparison of Fourier-Domain and Time-Domain Optical Coherence Tomography in the Detection of Band Atrophy of the Optic Nerve

被引:68
作者
Costa-Cunha, Luciana V. F. [1 ]
Cunha, Leonardo P. [1 ]
Malta, Roberto F. S. [1 ]
Monteiro, Mario L. R. [1 ]
机构
[1] Univ Sao Paulo, Sch Med, Hosp Clin, Neuroophthalmol Serv, Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
LAYER THICKNESS MEASUREMENTS; ULTRAHIGH-RESOLUTION; HIGH-SPEED; QUANTITATIVE-ANALYSIS; AXONAL LOSS; REPRODUCIBILITY;
D O I
10.1016/j.ajo.2008.07.020
中图分类号
R77 [眼科学];
学科分类号
100212 [眼科学];
摘要
PURPOSE: To compare the ability of Fourier-domain (FD) optical coherence tomography (3D OCT-1000; Top, con, Tokyo, Japan) and time domain (TD) OCT (Stratus; Carl Zeiss Meditec Inc, Dublin, California, USA) to detect axonal loss in eyes with band atrophy (BA) of the optic nerve. DESIGN: Cross-sectional study. METHODS: Thirty-six eyes from 36 patients with BA and temporal visual field (VF) defect from chiasmal compression and 36 normal eyes were studied. Subjects were submitted to standard automated perimetry and macular and retinal nerve fiber layer (RNFL) measurements were taken using 3D OCT-1000 and Stratus OCT. Receiver operating characteristic (ROC) curves were calculated for each parameter. Spearman correlation coefficients were obtained to evaluate the relationship between RNFL and macular thickness parameters and severity of VF loss. Measurements from the two devices were compared. RESULTS: Regardless of OCT device, all RNFL and macular thickness parameters were significantly lower in eyes with BA compared with normal eyes, but no statistically significant difference was found with regard to the area under the ROC curve. Structure-function relationships were also similar for the two devices. In both groups, RNFL and macular thickness measurements were generally and in some cases significantly smaller with 3D OCT-1000 than with Stratus OCT. CONCLUSIONS: The introduction of FD technology did not lead to better discrimination ability for detecting BA of the optic nerve compared with TD technology when using the software currently provided by the manufacturer. 3D OCT-1000 FD OCT RNFL and macular measurements were generally smaller than TD Stratus OCT measurements. Investigators should be aware of this fact when comparing measurements obtained with these two devices. (Am J Oplathalmol 2009;147: 56-63. (c) 2009 by Elsevier Inc. All rights reserved.)
引用
收藏
页码:56 / 63
页数:8
相关论文
共 20 条
[1]
Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[2]
Reproducibility of nerve fiber layer thickness measurements by use of optical coherence tomography [J].
Blumenthal, EZ ;
Williams, JM ;
Weinreb, RN ;
Girkin, CA ;
Berry, CC ;
Zangwill, LM .
OPHTHALMOLOGY, 2000, 107 (12) :2278-2282
[3]
Sensitivity advantage of swept source and Fourier domain optical coherence tomography [J].
Choma, MA ;
Sarunic, MV ;
Yang, CH ;
Izatt, JA .
OPTICS EXPRESS, 2003, 11 (18) :2183-2189
[4]
COMPARING THE AREAS UNDER 2 OR MORE CORRELATED RECEIVER OPERATING CHARACTERISTIC CURVES - A NONPARAMETRIC APPROACH [J].
DELONG, ER ;
DELONG, DM ;
CLARKEPEARSON, DI .
BIOMETRICS, 1988, 44 (03) :837-845
[5]
Peripapillary nerve fiber layer thickness profile determined with high speed, ultrahigh resolution optical coherence tomography high-density scanning [J].
Gabriele, Michelle L. ;
Ishikawa, Hiroshi ;
Wollstein, Gadi ;
Bilonick, Richard A. ;
Kagemann, Larry ;
Wojtkowski, Maciej ;
Srinivasan, Vivek J. ;
Fujimoto, James G. ;
Duker, Jay S. ;
Schuman, Joel S. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (07) :3154-3160
[6]
A framework for comparing structural and functional measures of glaucomatous damage [J].
Hood, Donald C. ;
Kardon, Randy H. .
PROGRESS IN RETINAL AND EYE RESEARCH, 2007, 26 (06) :688-710
[7]
OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[8]
Optical coherence tomography to detect and manage retinal disease and glaucoma [J].
Jaffe, GJ ;
Caprioli, J .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2004, 137 (01) :156-169
[9]
The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer [J].
Jones, AL ;
Sheen, NJL ;
North, RV ;
Morgan, JE .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (06) :673-677
[10]
Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucoma [J].
Medeiros, FA ;
Zangwill, LM ;
Bowd, C ;
Weinreb, RN .
ARCHIVES OF OPHTHALMOLOGY, 2004, 122 (06) :827-837