Are undesirable contact kinematics minimized after kinematically aligned total knee arthroplasty? An intersurgeon analysis of consecutive patients

被引:73
作者
Howell, Stephen M. [1 ]
Hodapp, Esther E. [1 ]
Vernace, Joseph V.
Hull, Maury L. [1 ]
Meade, Thomas D.
机构
[1] Univ Calif Davis, Dept Mech Engn, Davis, CA 95616 USA
关键词
Total knee arthroplasty; Kneeling; Kinematic alignment; Kinematics; ROTATIONAL ALIGNMENT; REPLACEMENT; COMPONENT;
D O I
10.1007/s00167-012-2220-2
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Tibiofemoral contact kinematics or knee implant motions have a direct influence on patient function and implant longevity and should be evaluated for any new alignment technique such as kinematically aligned total knee arthroplasty (TKA). Edge loading of the tibial liner and external rotation (reverse of normal) and adduction of the tibial component on the femoral component are undesirable contact kinematics that should be minimized. Accordingly, this study determined whether the overall prevalence of undesirable contact kinematics during standing, mid kneeling near 90 degrees and full kneeling with kinematically aligned TKA are minimal and not different between groups of consecutive patients treated by different surgeons. Three surgeons were asked to perform cemented, kinematically aligned TKA with patient-specific guides in a consecutive series of patients with their preferred cruciate-retaining (CR) implant. In vivo tibiofemoral contact positions were obtained using a 3- to 2-dimensional image registration technique in 69 subjects (Vanguard CR-TKA N = 22, and Triathlon CR-TKA N = 47). Anterior or posterior edge loading of the tibial liner was not observed. The overall prevalence of external rotation of the tibial component on the femoral component of 6 % was low and not different between surgeons (n.s.). The overall prevalence of adduction of the tibial component on the femoral component of 4 % was low and not different between surgeons (n.s.). Kinematically aligned TKA minimized the undesirable contact kinematics of edge loading of the tibial liner, and external rotation and adduction of the tibial component on the femoral component during standing and kneeling, which suggests an optimistic prognosis for durable long-term function. III.
引用
收藏
页码:2281 / 2287
页数:7
相关论文
共 29 条
[1]
The role of pain and function in determining patient satisfaction after total knee replacement - Data from the National Joint Registry for England and Wales [J].
Baker, P. N. ;
van der Meulen, J. H. ;
Lewsey, J. ;
Gregg, P. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (07) :893-900
[2]
Accurate measurement of three-dimensional knee replacement kinematics using single-plane fluoroscopy [J].
Banks, SA ;
Hodge, WA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (06) :638-649
[3]
Kneeling Is Safe for Patients Implanted With Medial-Pivot Total Knee Arthroplasty Designs [J].
Barnes, C. Lowry ;
Sharma, Adrija ;
Blaha, J. David ;
Nambu, Satya N. ;
Carroll, Michael E. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (04) :549-554
[4]
Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty - Influence of posterior condylar offset [J].
Bellemans, J ;
Banks, S ;
Victor, J ;
Vandenneucker, H ;
Moemans, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B :50-53
[5]
Malrotation causing patellofemoral complications after total knee arthroplasty [J].
Berger, RA ;
Crossett, LS ;
Jacobs, JJ ;
Rubash, HE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (356) :144-153
[6]
Kneeling kinematics after total knee arthroplasty: anterior-posterior contact position of a standard and a high-flex tibial insert design [J].
Coughlin, Kathryn M. ;
Incavo, Stephen J. ;
Doohen, Robert R. ;
Gamada, Kazuyoshi ;
Banks, Scott ;
Beynnon, Bruce D. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (02) :160-165
[7]
The 6 degrees of freedom kinematics of the knee after anterior cruciate ligament deficiency - An in vivo imaging analysis [J].
DeFrate, Louis E. ;
Papannagari, Ramprasad ;
Gill, Thomas J. ;
Moses, Jeremy M. ;
Pathare, Neil P. ;
Li, Guoan .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2006, 34 (08) :1240-1246
[8]
Kinematically Versus Mechanically Aligned Total Knee Arthroplasty [J].
Dossett, H. Gene ;
Swartz, George J. ;
Estrada, Nicolette A. ;
LeFevre, George W. ;
Kwasman, Bertram G. .
ORTHOPEDICS, 2012, 35 (02) :E160-E169
[9]
Kinematic analysis of kneeling in cruciate-retaining and posterior-stabilized total knee arthroplasties [J].
Hamai, Satoshi ;
Miura, Hiromasa ;
Higaki, Hidehiko ;
Matsuda, Shuichi ;
Shimoto, Takeshi ;
Sasaki, Kousuke ;
Yoshizumi, Masaaki ;
Okazaki, Ken ;
Tsukamoto, Nobuaki ;
Iwamoto, Yukihide .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2008, 26 (04) :435-442
[10]
Howell SM., 2012, INSALL SCOTT SURG KN, P1255