Risk analysis of Leksell Gamma Knife Model C with Automatic Positioning System

被引:17
作者
Goetsch, SJ [1 ]
机构
[1] San Diego Gamma Knife Ctr, La Jolla, CA 92037 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2002年 / 52卷 / 03期
关键词
Leksell Gamma Knife; quality assurance; medical errors; nosocomial infection; probable risk analysis;
D O I
10.1016/S0360-3016(01)02718-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: This study was conducted to evaluate the decrease in risk from misadministration of the new Leksell Gamma Knife Model C with Automatic Positioning System compared with previous models. Methods and Materials: Elekta Instruments, A.B. of Stockholm has introduced a new computer-controlled Leksell Gamma Knife Model C which uses motor-driven trunnions; to reposition the patient between isocenters (shots) without human intervention. Previous models required the operators to manually set coordinates from a printed list, permitting opportunities for coordinate transposition, incorrect helmet size, incorrect treatment times, missing shots, or repeated shots. Results: A risk analysis was conducted between craniotomy involving hospital admission and outpatient Gamma Knife radiosurgery. A report of the Institute of Medicine of the National Academies dated November 29, 1999 estimated that medical errors kill between 44,000 and 98,000 people each year in the United States. Another report from the National Nosocomial Infections Surveillance System estimates that 2.1 million nosocomial infections occur annually in the United States in acute care hospitals alone, with 31 million total admissions. Conclusions: All medical procedures have attendant risks of morbidity and possibly mortality. Each patient should be counseled as to the risk of adverse effects as well as the likelihood of good results for alternative treatment strategies. This paper seeks to fill a gap in the existing medical literature, which has a paucity of data involving risk estimates for stereotactic radiosurgery. (C) 2002 Elsevier Science Inc.
引用
收藏
页码:869 / 877
页数:9
相关论文
共 44 条
[1]  
[Anonymous], 2000, LANCET, V355, P2007
[2]   QUALITY ASSURANCE OF LEKSELL GAMMA-UNITS [J].
BERK, HW ;
AGARWAL, SK .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 1991, 57 (1-2) :106-112
[3]   INCIDENCE OF ADVERSE EVENTS AND NEGLIGENCE IN HOSPITALIZED-PATIENTS - RESULTS OF THE HARVARD MEDICAL-PRACTICE STUDY-I [J].
BRENNAN, TA ;
LEAPE, LL ;
LAIRD, NM ;
HEBERT, L ;
LOCALIO, AR ;
LAWTHERS, AG ;
NEWHOUSE, JP ;
WEILER, PC ;
HIATT, HH .
NEW ENGLAND JOURNAL OF MEDICINE, 1991, 324 (06) :370-376
[4]  
BREZOVICH IA, 1997, INT J RADIAT ONCOL, V38, P663
[5]   A RADIATION OVERDOSE INCIDENT - INITIAL DATA [J].
COHEN, L ;
SCHULTHEISS, TE ;
KENNAUGH, RC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (01) :217-224
[6]   ASSURANCE OF HIGH-QUALITY LINAC-BASED STEREOTAXIC RADIOSURGERY [J].
DRZYMALA, RE ;
KLEIN, EE ;
SIMPSON, JR ;
RICH, KM ;
WASSERMAN, TH ;
PURDY, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 30 (02) :459-472
[7]   Use of a micro-ionization chamber and an anthropomorphic head phantom in a quality assurance program for stereotactic radiosurgery [J].
Duggan, DM ;
Coffey, CW .
MEDICAL PHYSICS, 1996, 23 (04) :513-516
[8]   Quality assurance for the Leksell gamma unit: Considering magnetic resonance image-distortion and delineation failure in the targeting of the internal auditory canal [J].
Ertl, A ;
Saringer, W ;
Heimberger, K ;
Kindl, P .
MEDICAL PHYSICS, 1999, 26 (02) :166-170
[9]   POTENTIAL HUMAN ERROR IN SETTING STEREOTAXIC COORDINATES FOR RADIOSURGERY - IMPLICATIONS FOR QUALITY ASSURANCE [J].
FLICKINGER, JC ;
LUNSFORD, LD ;
KONDZIOLKA, D ;
MAITZ, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02) :397-401
[10]   American Association of Physicists in Medicine radiation therapy committee task group 53: Quality assurance for clinical radiotherapy treatment planning [J].
Fraass, B ;
Doppke, K ;
Hunt, M ;
Kutcher, G ;
Starkschall, G ;
Stern, R ;
Van Dyke, J .
MEDICAL PHYSICS, 1998, 25 (10) :1773-1829